SERVICE MANUAL
Invacare
®
- Kite
22
Component
Check
Remedy
Notes
ü
Legrests
•
Check welded
seams, interlocking,
screws, footplates
•
Tighten, replace if
necessary
Power legrests
(if fitted)
•
Check cable
•
Check contacts
•
check functions
•
Replace cable if
necessary
Lighting (if
fitted)
•
Check cable
•
Check function
•
Replace lamp or
cable if necessary
Batteries
•
Check batteries for
damage
•
Replace batteries
if necessary
See chapter 8.4.
•
Check battery
voltage
•
Charge batteries
See operating
manual
•
Check contacts and
terminals
•
Clean contacts
and terminals
Please refer to the
safety information
in Chapter 8.4 for
handling batteries
Remote /
electronic
module
•
Remote, status
display blinking
•
Evaluate
error/blink code
•
Fixings
•
Tighten fixings,
replace if
necessary
•
Cables and
connecting plugs
•
Tighten cables
and connecting
plugs, replace if
necessary
•
Drive lever function
•
Replace drive
lever if necessary
•
Replace remote if
necessary
•
Power supply
•
Tighten cables
and connecting
plugs, replace if
necessary
Drive program
•
Check drive
electronics program
version
•
Update software if
newer version
available.