SERVICE MANUAL
Invacare
®
- Kite
90
•
Remove the SL fuse (2) on the bottom fixing
bolt (1) with the screwdriver.
•
Pull the fixing bolt (1) out.
•
Remove the SL fuse (2) on the top fixing bolt
(1) with the screwdriver.
•
Pull the fixing bolt (1) out.
•
Pull the adjustment spindle out together with the actuator.
•
Insert the new adjustment spindle with the actuator.
•
Reinsert the fixing bolts and SL fuses. Fix the SL fuses securely to the fixing bolt with the
hammer.
•
Connect the plug of the actuator to the electronic module.
•
Fasten the actuator cable to the frame with new cable ties. Position the cable such that it
does not get abraded or pinched.
•
To complete, check all seat tilting functions.