SERVICE MANUAL
Invacare
®
- Kite
122
9.3.2
Replacing the connecting cable of the operating hours counter
Requirements:
•
Philipps screwdriver, size 2
•
Side cutters
•
Cable ties
•
The operating hours counter is located in the
side cover.
•
Disconnect the plug-in connection of the
operating hours counter from the drive
motor.
•
Remove the side cover, as described in
chapter 8.1.3.
•
Detach the operating hours counter from the
side cover.
•
Turn the operating hours counter around.
•
Loosen and remove the four Phillips screws
of the housing.
•
Disconnect the plugs (+) & (-).
•
Remove the cable ties (4).
•
Carefully pull the connecting cable (2) of the
operating hours counter out of the casing
(1).
•
Replace the cable of the operating hours
counter.