18 . Electro Magnetic Compliance (EMC)
The Medical Bed is intended for use in the electromagnetic environment specified below. The user of the Medical
bed should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Medical Bed uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Medical Bed is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Medical Bed is intended for use in the electromagnetic environment specified below. The user of the Medical
should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601
test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output
lines
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
± 1 kV differential
mode
± 2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
Medical bed Alegio requires
continued operation during power
mains interruptions, it is recommended
that the Medical Bed Alegio be
powered from an uninterruptible
power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
0,3 A/m
The power frequency magnetic field
should be at a characteristic level of a
typical commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
24
EN.indd 24
18/07/11 18:27
Содержание Alegio Series
Страница 4: ......
Страница 11: ...D C F E G H EN 13 ...
Страница 31: ...D C F E G H SV 13 ...
Страница 51: ...D C F E G H NL 13 ...
Страница 71: ...D C F E G H FR 13 ...
Страница 91: ...D C F E G H ES 13 ...
Страница 111: ...D C F E G H IT 13 ...
Страница 131: ...D C F E G H PT 13 ...
Страница 145: ......
Страница 146: ......
Страница 147: ......