background image

120-00012  
GasD 8000 Series User Manual

 

12/14/2020

 

 

      2.1.2   CONTROL KEYPAD 

The analyzer

’s control keypad is detailed in Figure 4 below. Each button is described 

in the table that follows. 

 

 

 

 

 

 

 

 

 

 

 

 

 

POWER BUTTON 

– 

Press to turn power 

to  instrument  ON.  Press  from  the  MAIN 
MENU to turn power OFF. 
 

ARROW  BUTTONS 

– 

Used  to  navigate 

through the instrument’s menus and enter 
values  during  numeric  and  text  entry. 
Proper use is indicated in specific function 
descriptions. 
 
 

ENTER  BUTTON 

– 

Used  to  advance 

through the sequential menus and finalize 
certain functions. 
 

STOP  BUTTON 

– 

Used  to  back  up  in 

sequential  menus  or  end  a  particular 
function. 
 

SAVE  BUTTON 

– 

Press  at  any  point 

during  sampling  to  save  current  data  to 
memory card. 
 

 

 

2.3    SAMPLE PROBE 

 

 

In most cases, gas sample will be collected via a provided SAMPLE PROBE which consists 

of probe handle and tip, a 3 foot length of PTFE (Teflon) or polypropylene tubing and a 

quick  disconnect  tube  adapater  for  connection  to  the  INLET  of  the  instrument.  This 

assembly is shown below.  

POWER BUTTON 

ARROW BUTTONS 

ENTER BUTTON 

STOP BUTTON 

SAVE BUTTON 

FIGURE 4 

Содержание 120-00012

Страница 1: ...120 00012 GasD 8000 Series User Manual 12 14 2020 I Interscan Corporation Instruction Manual GasD 8000SeriesPortableGasAnalyzer P N 120 00012...

Страница 2: ...Alarm Indications 13 3 7 Charging The Batteries 14 Section 4 Sampling 16 4 1 Zeroing The analyzer 16 4 2 Taking Samples 16 4 3 Sample Interferences 17 4 4 Auto Blockage Detection 17 Section 5 Data Sto...

Страница 3: ...ible and display indications as well as voltage output at rear panel terminals Voltage and current loop analog outputs available at rear panel terminals Rechargeable Li ion battery providing 5 8 hours...

Страница 4: ...Be sure to read section 4 3 on Sample Interferences carefully before taking any gas samples Carefully remove the analyzer from its packing container along with the accessories Inspect the instrument f...

Страница 5: ...2 for connection details OUT FITTING Exhaust port for sample stream No connection is necessary DO NOT BLOCK THIS PORT MINI USB PORT USB connector for battery charging and powered operation See sectio...

Страница 6: ...nt of the instrument It will lock into place when at the 180 position Be sure to collapse the tilt back feet before rotating the handle FIGURE 2 2 1 1 DISPLAY FIELDS AND INDICATORS The GasD 8000 serie...

Страница 7: ...descriptions ENTER BUTTON Used to advance through the sequential menus and finalize certain functions STOP BUTTON Used to back up in sequential menus or end a particular function SAVE BUTTON Press at...

Страница 8: ...while pulling out on the tubing below right 2 4 ANALOG OUTPUT CONNECTIONS The GasD 8000 Series analyzer is equipped with 0 100 mV and a 4 20 mA analog output signals available at the 10 pin rear panel...

Страница 9: ...ence N C N C Alarm Signal Analog Voltage Output Analog Current Output N C N C Local Device Ground Local Device Ground 5 Volts 5V on alarm actuation 0 100 mV corresponding to gas range 4 20 mA correspo...

Страница 10: ...MAIN MENU screen shown on the right and press the green power button The screen will prompt you to confirm power down as shown Press the RIGHT ARROW button to confirm power down NOTE Normal power dow...

Страница 11: ...PLING MODE DATE TIME ALARM GAS CALIBRATION DIGITAL CALIBRATION FACTORY SETTINGS See section 3 6 for details on the SETUP sub menu FILES Sub menu of data storage parameters as follows Create create and...

Страница 12: ...MODE the pump is turned off and the screen showed on the right will be displayed The timer at the bottom of the display indicates how long the unit has been in an operational mode This is the mode th...

Страница 13: ...enu shown on the right offers access to user adjustable parameters and maintenance functions This menu is detailed below Flow rate cc min Nominal setting 200 cc min unless stated otherwise Do not chan...

Страница 14: ...r factory setup only The GasD 8000 Series portable analyzer is equipped with an alarm feature that provides visual and audible indication when the gas concentration level exceeds a user set alarm leve...

Страница 15: ...ter value field NOTE Decimal places displayed in the Alarm numeric value field will mirror the decimal place count of the primary gas range Use the DOWN ARROW button to highlight the ENABLE parameter...

Страница 16: ...ill indicate Low Battery at the bottom of the screen Follow the procedure below to re charge the batteries 1 Select the desired plug type on the multi region charger plug unit 2 Plug the large USB con...

Страница 17: ...do not connect any other USB devices to the charger NOTE The internal batteries must have a charge to maintain the sensor bias voltage and minimize sensor warmup time It is recommended you keep the i...

Страница 18: ...hlighted indicating ZERO MODE is active Press the ENTER button Set ZERO will be indicated under the main display for 3 seconds indicating the value is being zeroed The display value will gradually dec...

Страница 19: ...discussed in this Knowledge Base article https www gasdetection com gas detection knowledge base ethylene oxide eto or eo monitoring applications alcohol and electrochemical sensors For further infor...

Страница 20: ...in AUTO SAVE mode is the surest way to ensure data storage is completed The trade off is the potential for storing long strings of unnecessary data every second in ON mode For this reason the option...

Страница 21: ...t the desired file in which to store the data for the next sample cycle s Highlight the desired file using the UP DOWN arrows then press the RIGHT ARROW button to select The currently selected file wi...

Страница 22: ...he SD card slot NOTE If the SD card is not inserted into the instrument s SD card slot on power up NO SD Card will be displayed in the lower left corner of the LCD display and AUTO SAVE will be disabl...

Страница 23: ...aken from a gas cylinder or a permeation device Besides being essential for calibration having a known certified gas standard on hand allows the user to test the analyzer at any time to confirm proper...

Страница 24: ...press the RIGHT ARROW button to enter the SETUP MENU 3 Use the DOWN ARROW button to move the cursor and highlight Gas Calibration as shown to the right and press the RIGHT ARROW button to enter the G...

Страница 25: ...e A 2 minute Calibration timer will begin counting down as indicated on the right while Calibration gas is drawn through the sensor The Reading indication shows the real time sensor response which can...

Страница 26: ...certificate detailing its sensitivity data that is entered during the Digital Calibration procedure detailed in the section 6 4 below The sensor replacement procedure is detailed in section 7 6 4 DIGI...

Страница 27: ...value shown matches the value on the ECS certification When satisfied with the entry press the LEFT ARROW button successively until the Main Menu screen is displayed 5 Turn power to the unit OFF by pr...

Страница 28: ...the rear bezel away from the side of the instrument case while pulling the bezel toward you The bezel will pivot away from the case as shown in figure 7 2 3 Slowly continue pivoting the right side of...

Страница 29: ...The sensor is held in a manifold that is attached to the left side of the instrument case looking from the rear as shown below in Figure 7 7 7 Reaching carefully into the instrument case grasp the bl...

Страница 30: ...manifold The senor should release from the manifold as shown in figure 7 11 1 Press the new sensor firmly into the manifold until you feel it snap into place Orient with the terminal pins positioned...

Страница 31: ...the screw out and re align the panel holes with the threaded holes Forcing a misaligned screw into the threaded hole will strip the threads 3 Re connect the terminal block to the connector socket ter...

Страница 32: ...alibration Unit will not power up Batteries may need charging Use the charger cable to connect the wall charger to the charging port on the unit The display should indicate charging in progress Eject...

Страница 33: ...epaired or altered by unauthorized persons or which have been subject to misuse negligence or accident incorrect wiring by others installation or use not in accordance with instructions furnished by t...

Страница 34: ...the TROUBLESHOOTING section of this manual section 7 page 23 as this may help you to resolve any problems without returning the unit Should consulting the TROUBLESHOOTING section of the manual not add...

Отзывы: