120-00012
GasD 8000 Series User Manual
12/14/2020
23
6) On the Sensitivity Input screen, enter the
value of the Calibration Gas being used in the
field indicated on the right. Use the
RIGHT
and
LEFT ARROW
buttons to move the
cursor to the desired digit and the
UP
and
DOWN ARROW
buttons to change the value
as needed. Double check that the value is
accurate then press the
ENTER
button to proceed.
7) Connect the inlet to the CAL GAS source. A
2-minute Calibration timer will begin counting
down as indicated on the right while
Calibration gas is drawn through the sensor.
The “Reading” indication shows the real time
sensor response which can be compared to
the “Cal Gas” value.
8)
When the Calibration timer fully elapses, the
“Confirm Calibration” screen shown on the
right will be displayed.
The “Cal Reading” field
indicates the final reading at the end of the
Calibration Timer period. The “Cal Gas” value
is shown below the reading value and the
final Sensitivity value is shown below that.
Press the
SAVE
button to confirm calibration or press the
STOP
button to cancel
calibration.