120-00012
GasD 8000 Series User Manual
12/14/2020
24
There are circumstances in which calibration using calibration gas or permeation devices
is inconvenient and costly. For this reason, INTERSCAN has developed the alternative
sensor replacement program,
Sensor Express
.
With
Sensor Express
, a certified spare senso
r is kept at the customer’s site to be installed
into the analyzer, replacing the older sensor. The sensor being replaced is discarded. The
new sensor is accompanied by a certificate detailing its sensitivity data that is entered
during the
Digital Calibration
procedure detailed in the section 6.4 below. The sensor
replacement procedure is detailed in section 7.
6.4 DIGITAL CALIBRATION PROCEDURE
1) Navigate to the MAIN MENU by pressing the
STOP
button successively until the MAIN
MENU shown to the right is displayed.
2) Press the
UP or
DOWN ARROW
buttons to
move the cursor to highlight “
Setup
” then
press the
RIGHT ARROW
button to enter the
SETUP MENU.
3) Use the
DOWN ARROW
button to move the
cursor and highlight “
Digital Calibration
” as
shown to the right and press the
RIGHT
ARROW
button to enter the Digital Calibration
Menu.
6.3 DIGITAL CALIBRATION