120-00012
GasD 8000 Series User Manual
12/14/2020
22
Once the Calibration Gas delivery method has been determined, follow the procedure
below to carry out Gas Calibration:
1) Navigate to the MAIN MENU by pressing the
STOP
button successively until the MAIN
MENU shown to the right is displayed.
2) Press the
UP or
DOWN ARROW
buttons to
move the cursor to highlight
“
Setup
” then
press the
RIGHT ARROW
button to enter the
SETUP MENU.
3) Use the
DOWN ARROW
button to move the
cursor and highlight
“
Gas Calibration
” as shown
to the right and press the
RIGHT ARROW
button
to enter the Gas Calibration Menu.
4) You will now be prompted to carry out zeroing as
shown to the right. With the analyzer
drawing
clean ambient
air, press the
ENTER
button to
initiate zeroing.
5) A 2-minute Zero timer will begin counting
down as shown on the right after which the
unit will automatically zero itself and jump to
the Sensitivity input screen.
6.2 GAS CALIBRATION PROCEDURE