MAINTENANCE PLAN
Synthetic leather
Clean synthetic leather covers with a cloth and com-
mercially available mild soap solution�
Spacer fabrics/mesh material
Wash removable covers made of spacer fabrics/mesh
material at a maximum of
30
°C in the washing machine
with the delicates program�
Supra-tex
interco uses Supra-tex for the arm rests� Clean these
regularly with a cloth and commercially available mild
soapy water�
Neoprene
Wash removable fixation aids made of neoprene at a
maximum of
30
°C in the washing machine� Use a mild
liquid detergent or a commercially available neoprene
detergent� Use a gentle wash cycle and do not spin-dry
the neoprene covers� Then hang up the covers wet to
dry�
Trevira CS (fabrics)
Removable covers made of Trevira CS can be dry
cleaned� Washing under easy-care conditions at
60
°C
is possible� Do not dry and do not iron!
Dartex
Test washed by the manufacturer at temperatures
between
40
and
95
°C�
Accessory components made of metal, plastic, wood
Please clean accessory components with a conven-
tional mild household soap solution�
11
.
1
.
3
. Disinfection of the product
Please use a water-based agent for disinfection of
the product, e�g� Terralin, Quartamon Med or Sagro-
tan Original concentrate� Please observe the manu-
facturer's instructions for use when doing so� When
disinfecting the fabrics, also follow the instructions
of the fabric manufacturer regarding the use of brand
name disinfectants�
Rinse off all detergents and disinfectants thoroughly
and leave the cleaned object to dry completely before
use� If this is neglected, this can lead to chemical reac-
tions making the material brittle or lead to skin irrita-
tions of the patient�
Important!
No disinfectants for synthetic leather/skai
Synthetic leather covers/skai covers
may not be cleaned with disinfecting agents, as
the material gets porous over time and cracks�
Note!
Wipe disinfection
If wipe disinfection is necessary, use a
disinfectant that is suitable for surface disinfec-
tion according to the DGH disinfectant list�
11
.
2
. Maintenance
Your interco product is checked by trained and qualified
employees for proper functioning and freedom from
defects before being handed over to the user� To
ensure safe use of the product over its entire service
life (according to the manufacturer from min�
2
years
to max�
6
years) and to ensure a perfect technical
condition, it must be maintained (every
6
months) and
inspected (every
12
month) regularly according to the
following maintenance and service plans�
Note!
Changing settings
The changing of settings (other than
those described in Section
7
) is only permitted for
qualified and trained personnel!
Incorrect settings by the user, operator or laypersons
bear the risk of injuries or crushing, which in a worst
case scenario could lead to the patient's death!
During normal use, you should regularly check the
interco product for damage and the perfect working
condition of all components�
Important!
Original spare parts
Only original spare parts may be used
for service and maintenance work
Содержание SKIPPY 62500-1
Страница 1: ...08 2020 Gebrauchsanweisung Operating instructions SKIPPY...
Страница 2: ...Gebrauchsanweisung SKIPPY Operating instructions SKIPPY Deutsch ab Seite 3 English starting on page 29...
Страница 3: ...3 Deutsch Gebrauchsanweisung SKIPPY Deutsch...
Страница 26: ...26 EU KONFORMIT TSERKL RUNG 15 EU Konformit tserkl rung...
Страница 28: ......
Страница 29: ...29 English Operating instructions SKIPPY English...
Страница 52: ...52 EU DECLARATION OF CONFORMITY 15 EU Declaration of Conformity...
Страница 54: ...Notizen Notes...
Страница 55: ......