PS 9000 1U Series  •  DC Laboratory Power Supply  •  User Manual

4.3 

Calibration (readjustment)

4.3.1 Preface

The devices of series PS 9000 1U feature a function to readjust the most important output values when doing 

a calibration and in case these values have moved out of tolerance. The readjustment is limited to compensate 

small differences of up to 1% or 2% of the max. value. There are several reasons which could make it necessary 

to readjust a unit: component aging, component deterioration, extreme ambient conditions, high frequent use
In order to determine if a value is out of tolerance, the parameter must be verified first with measurement tools of 

high accuracy and with at least half the error of the PS device. Only then a comparison between values displayed 

on the PS device and true DC output values is possible.
For example, if you want to verify and possibly readjust the output current of model PS 9080-100 1U which has 

100 A maximum current, stated with a max. error of 0.2%, you can only do that by using a high current shunt with 

max. 0.1% error or less. Also, when measuring such high currents, it is recommended to keep the process short, 

in order to avoid the shunt heating up too much. It is furthermore recommended to use a shunt with at least 25% 

reserve.
When measuring the current with a shunt, the measurement error of the multimeter on the shunt adds to the error 

of the shunt and the sum of both must not exceed the max. error of the device under calibration.

4.3.1.1 Preparation

For a successful calibration and readjustment, a few tools and certain ambient conditions are required:

• 

A measurement device (multimeter) for voltage, with a max. error of half the PS’s voltage error. That measure-

ment device can also be used to measure the shunt voltage when readjusting the current

• 

If the current is also going to be calibrated: a suitable DC current shunt, ideally specified for at least 1.25 times 

the max. output current of the PS and with a max. error that is half or less than the max. current error of the PS 

device

• 

Normal ambient temperature of approx. 20-25°C

• 

Warmed up PS unit, which has been run for at least 10 minutes under 50% power

• 

An adjustable load, such as an electronic load, which is capable of consuming at least 102% of the max. voltage 

and current of the PS device

Before you can start calibrating, a few measures have to be taken:

• 

Let the PS device warm up in connection with the voltage / current source

• 

In case the remote sensing input is going to be calibrated, prepare a cable for the remote sensing connector to 

DC output, but leave it yet unconnected

• 

Abort any form of remote control, deactivate master-slave mode, set device to 

U/I

 mode

• 

Install the shunt between PS device and load and make sure the shunt is cooled somehow

• 

Connect external measurement device to the DC output or to the shunt, depending on whether the voltage is 

going to be calibrated first or the current

4.3.2  Calibration procedure

After the preparation, the device is ready to be calibrated. From now on, a certain sequence of parameter calibra-

tion is important. Generally, you don’t need to calibrate all three parameters, but it is recommended to do so.
Important:

When calibrating the output voltage, the remote input “Sense” on the rear of the device has to 

be disconnected.

The calibration procedure, as explained below, is an example with model PS 9080-100 1U. Other models are 

treated the same way, with values according to the particular PS model and the required load.

Содержание PS 9080-100 1U

Страница 1: ...THE POWER TEST EXPERTS User Manual Attention This document is only valid for devices with firmware KE 2 07 and HMI 2 03 or higher For availability of updates for your device check our website or cont...

Страница 2: ......

Страница 3: ...ting the USB port 26 2 3 11 Initial commission 26 2 3 12 Initial network setup 27 2 3 13 Commission after a firmware update or a long period of non use 27 3 OPERATIONANDAPPLICATION 3 1 Important notes...

Страница 4: ...tepro Systems LP Specifications subject to change without notice 4 3 Calibration readjustment 51 4 3 1 Preface 51 4 3 1 1 Preparation 51 4 3 2 Calibration procedure 51 5 CONTACTAND SUPPORT 5 1 General...

Страница 5: ...document 1 1 4 Explanation of symbols Warning and safety notices as well as general notices in this document are shown in a box with a symbol as follows Symbol for a life threatening danger Symbol for...

Страница 6: ...of the device in Volt Series 8 Series 8000 or 800 9 Series 9000 Type identification PS Power Supply usually programmable PSI Power Supply Intelligent always programmable ELR Electronic Load with Reco...

Страница 7: ...equipment and to the source ESD regulations must be applied when plugging interface cards or modules into the relative slot Interface cards or modules may only be attached or removed after the device...

Страница 8: ...cleaning of the equipment ensure that all employees who use the equipment have read and understood the manual Furthermore the users are to be regularly schooled in working with the equipment and the...

Страница 9: ...hold was lower than the actual output voltage Critical The device and or the load could be damaged Signal OCP OverCurrent Shutdown of the DC output due to excess of the preset limit Non critical prote...

Страница 10: ...1 7 ms 1 4 ms 2 2 ms 2 ms 2 ms Display Resolution See section 1 9 4 4 Resolution of the displayed values Display Accuracy 3 0 2 UNom 0 2 UNom 0 2 UNom 0 2 UNom 0 2 UNom Ripple 2 100 mVPP 5 2 mVRMS 29...

Страница 11: ...1150 V DC Miscellaneous Cooling Temperature controlled fans front inlet rear exhaust Ambient temperature 0 50 C Storage temperature 20 70 C Humidity 80 not condensing Standards EN 61010 EN 61326 Overv...

Страница 12: ...s 2 2 ms 2 ms 2 ms Display Resolution See section 1 9 4 4 Resolution of the displayed values Display Accuracy 3 0 2 UNom 0 2 UNom 0 2 UNom Ripple 2 76 mVPP 4 2 mVRMS 234 mVPP 40 mVRMS 156 mVPP 26 mVRM...

Страница 13: ...V DC 1150 V DC Miscellaneous Cooling Temperature controlled fans front inlet rear exhaust Ambient temperature 0 50 C Storage temperature 20 70 C Humidity 80 not condensing Standards EN 61010 EN 61326...

Страница 14: ...cifications subject to change without notice Figure 1 Front side Figure 2 Rear side 1 8 4 Views A Mains switch D Exhausts G AC input connector B Control panel E Share Bus and remote sensing connectors...

Страница 15: ...15 PS 9000 1U Series PS 9000 1U Series DC Laboratory Power Supply User Manual Figure 3 Left hand side view with DC cover Figure 4 Right hand side view without DC cover...

Страница 16: ...16 PS 9000 1U Series 2015 Intepro Systems LP Specifications subject to change without notice Figure 5 Top view...

Страница 17: ...ubmenus and parameters direction up left or changes view mode Button Navigates through menus submenus and parameters direction down right or changes view mode Button Submits altered parameters or set...

Страница 18: ...to 1500 V DC In addition the devices offer as standard the possibility for parallel connection in Share bus operation for constant current sharing Operating in this way allows up to 10 units to be com...

Страница 19: ...pushing the corresponding rotary knob in order to move the cursor The left hand rotary knob is always assigned to the DC output voltage and related parameters like OVP or U max whereas the right hand...

Страница 20: ...ts The number of decimal places depends on the device model The values have 4 or 5 digits while actual and set values related to the same physical unit always have the same number of digits Adjustment...

Страница 21: ...other reserved ports The standard port for this device is 5025 default IP is 192 168 0 2 Via TCP IP and this port communication to the device can be established in most of the common programming lang...

Страница 22: ...ore commissioning the equipment should be visually inspected for damage and completeness using the delivery note and or parts list see section 1 9 3 Scope of delivery An obviously damaged device e g l...

Страница 23: ...et or multi socket outlet as long as the socket features a safety contact PE and is capable for 16 A When connecting the device to a multi socket outlet along with other electric devices it is importa...

Страница 24: ...erage ambient temperature up to 50 C we recommend up to 10 A 0 75 mm AWG18 up to 15 A 1 5 mm AWG14 up to 30 A 4 mm AWG10 up to 40 A 6 mm AWG8 up to 60 A 16 mm AWG4 up to 100 A 25 mm AWG2 per lead mult...

Страница 25: ...ad The device recognizes the remote sensing mode automati cally and regulates the output voltage only in CV operation at the load rather than at its own DC output In the technical specifications see s...

Страница 26: ...it is strongly recommended to use and install the included driver installer on the USB stick to gain maximum compatibility of the device to our software 2 3 10 2 Driver installation Linux MacOS We ca...

Страница 27: ...P http 192 168 0 2 or the default host name http Client only possible if there is a running DNS in the network into the URL box 3 After the website has been completely loaded check the status field it...

Страница 28: ...output power according to P UOUT IOUT reaches the adjusted current or power limit In both cases the device will automatically change to constant current or constant power operation whichever occurs fi...

Страница 29: ...current according to I P U The adjustable set value of current as indicated in the display is always a upper limit While the DC output is switched on and constant power mode is active then the conditi...

Страница 30: ...witch on again to continue operation 3 3 3 Overvoltage An overvoltage alarm OVP will switch off the DC output and can occur if the power supply itself as a voltage source generates an output voltage h...

Страница 31: ...red 3 4 2 Switching off the device On switch off the last output condition and the most recent set values are saved Furthermore a PF alarm power failure will be reported but can be ignored The DC outp...

Страница 32: ...min ____A I max ____A P max ____W General Calibration Reset Communication IP Settings 2 HMI Setup Language English Deutsch Display Language Brightness 1 10 HMI Lock HMI lock Lock all ON OFF possible E...

Страница 33: ...o switch off DC output after PF alarm 6 Determines how the DC output shall react after a power fail PF alarm has occurred OFF DC output will be switched off and remain until user action AUTO DC output...

Страница 34: ...IP address in standard IP format Subnet mask Only available with setting Get IP address Manual Default value 255 255 255 0 Permanent manual setting of the subnet mask in standard IP format Gateway Onl...

Страница 35: ...e That also includes remote control via analog or digital interface In remote control the global range of 0 100 digital resp 0 5 V 0 10 V remains only narrowed by the limits defined here An example yo...

Страница 36: ...after leaving the setup menu and the right hand rotary knob is then either assigned to set value of current or power 3 4 6 Manual adjustment of set values Adjusting set values of voltage current and p...

Страница 37: ...tput on during manual operation or digital remote control can be disabled by pin REM SB of the built in analog interface For more information refer to 3 4 3 1 and example a in 3 5 4 6 How to manually...

Страница 38: ...via the digital interface is active Remote then remote control is immediately terminated and must be reactivated at the PC once Local is no longer active If remote control is via the analog interface...

Страница 39: ...value outputs CMON VMON Input of excess set values e g 5 V in selected 5 V range or 10 V in the 10 V range are clipped by setting the set value at 100 Before you begin please read Important notes for...

Страница 40: ...5 V correspond to 0 100 von PNom Accuracy 0 5 V range 0 4 Accuracy 0 10 V range 0 2 Input im pedance Ri 40 k 100 k 9 VMON AO Actual voltage 0 10 V or 0 5 V correspond to 0 100 von UNom Accuracy 0 2 at...

Страница 41: ...n 2 03 this is also available without remote control being active It is recommended that a low resistance contact such as a switch relay or transistor is used to switch the pin to ground DGND The foll...

Страница 42: ...ated Pin Remote LOW The set values VSEL and CSEL are generated from the reference voltage VREF using potentiometers for each Hence the power supply can selectively work in current limiting or power li...

Страница 43: ...in REM SB to the level that corresponds to DC output off then switch it on again See section 3 5 4 6 Application examples for levels and logic How to acknowledge an alarm in the alarm buffer status du...

Страница 44: ...between a simple lock and a PIN lock which requires to enter the correct PIN every time someone wants to unlock the HMI How to lock the HMI 1 Switch off the DC output and push button to call the setup...

Страница 45: ...nge in HMI language On calling up the menu page and selecting a profile the most important settings can be seen but not changed How to save the current values and settings work profile as a user profi...

Страница 46: ...cted to the next unit using cables with cross section according to the maximum current and with short as possible length 3 9 1 2 Wiring the Share bus The Share bus is wired from unit to unit with an i...

Страница 47: ...same kind and model i e power supply with power supply like for example PS 9080 100 1U with PS 9080 100 1U or similar models like PS PSI 9080 120 2U Series connection is not supported by the software...

Страница 48: ...ed limits for the application should be set in the power supply This can be done via the interface The electronic load should preferably be in operating mode CV The load will then using the Share Bus...

Страница 49: ...put s of all units switched off In this combination of devices it is recommended to always switch on the DC output of the source first and then the DC input of the sink 1 Discharge of the battery to 2...

Страница 50: ...ipping destination is abroad the necessary customs papers are attached 4 2 1 Replacing a defect mains fuse The device is protected by one fuse T16 A 250 V 35 mm which is located inside the device This...

Страница 51: ...imeter for voltage with a max error of half the PS s voltage error That measure ment device can also be used to measure the shunt voltage when readjusting the current If the current is also going to b...

Страница 52: ...and start to measure the output current I mon Simply confirm here with 4 After this the device requests you to enter the output current which you are measuring with the shunt at Measured value Enter i...

Страница 53: ...4 3 2 2 Actual values Actual values of output voltage with and without remote sensing and output current are calibrated almost the same way as the set values but here you don t need to enter anything...

Страница 54: ...86 sales inteproate com www inteproate com United Kingdom Intepro UK Ltd 9 Lakeside Business Park Swan Lane Sandhurst Berkshire GU47 9DN UK Tel 44 012 5287 5600 China Intepro Power Block 7 Fourth Indu...

Отзывы: