PS 9000 1U Series
PS 9000 1U Series • DC Laboratory Power Supply • User Manual
1500 W
Model 1U
PS 9080-50
PS 9200-25
PS 9360-15
PS 9500-10
PS 9750-06
Analog interface
(1
Type
Sub-D, 15 pole, female
Set value inputs
U, I, P
U, I, P
U, I, P
U, I, P
U, I, P
Actual value output
U, I
U, I
U, I
U, I
U, I
Control signals
DC on/off,
Remote on/off
DC on/off,
Remote on/off
DC on/off,
Remote on/off
DC on/off,
Remote on/off
DC on/off,
Remote on/off
Status signals
CV, OVP, OT
CV, OVP, OT
CV, OVP, OT
CV, OVP, OT
CV, OVP, OT
Galvanic isolation to the device
Max. 1500 V DC Max. 1500 V DC Max. 1500 V DC Max. 1500 V DC Max. 1500 V DC
Insulation
Allowed float (potential shift) on the DC output:
Negative terminal to PE
Max. ±400 V DC
±400 V DC
±400 V DC
±400 V DC
±400 V DC
Positive terminal to PE
Max. +480 V DC
+600 V DC
+760 V DC
+900 V DC
+1150 V DC
Miscellaneous
Cooling
Temperature controlled fans, front inlet, rear exhaust
Ambient temperature
0..50°C
Storage temperature
-20...70°C
Humidity
< 80%, not condensing
Standards
EN 61010, EN 61326
Overvoltage category
2
Protection class
1
Pollution degree
2
Operational altitude
< 2000 m
Digital interfaces
Featured
1x USB-B, 1x Ethernet
Galvanic isolation from device
Max. 1500 V DC
USB specification
USB 2.0, socket type B, VCOM driver
USB response time
SCPI: max. 5 ms, ModBus: max. 5 ms
Ethernet specification
RJ45, 10/100Mbit, TCP/IP, ICMP, HTTP, DHCP
Ethernet response time
SCPI: max. 7 ms, ModBus: 9-17 ms
Terminals
Rear side
Share Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD)
19” x 1U x 463 mm
Total (WxHxD)
483 x 44 x min. 565 mm
Weight
~ 10.5 kg
~ 10.5 kg
~ 10.5 kg
~ 10.5 kg
~ 10.5 kg
Article number
06230400
06230401
06230402
06230403
06230404
(1 For technical specifications of the analog interface see
“3.5.4.3 Analog interface specification” on page 40