PS 9000 1U Series

© 2015 Intepro Systems, LP. Specifications subject to change without notice.

3.4.5  Display modes for actual and set values

The display of a PS 9000 1U device shows the actual output voltage and the related set value in the left half of the 

display and the actual output current and related set value in the right half. To have the power set value in direct 

access, the display mode can be switched.

Mode 

UI

Only voltage (U) and current (I) are displayed. This is the default mode.

The set value of power is then only accessible in the setup menu or 

when switching to a different view mode with the arrow buttons 

 

or 

.

Mode 

UP

Alternatively to the actual and set values of current (I) and voltage (U) 

the actual and set values of power (P) are displayed.

The set value of current is then only accessible in the setup menu or 

when switching to a different view mode with the arrow buttons 

 

or 

.

 

How to change the display mode in the menu

1. 

Switch off the DC output and push button 

 to call the setup menu.

2. 

In the menu navigate to “

HMI Setup”

 and push 

. Then in the submenu navigate to “

Page Setup

” and 

push 

 again.

3. 

Select the desired display mode (see above) with the right-hand rotary knob.

4. 

Accept the settings with 

 or discard them with 

.

 

How to change the display mode directly

1. 

During normal display (as shown in the figures above), push any of the arrow buttons 

 

 to switch 

between the view modes.

Depending on your setting, the main display will change after leaving the setup menu and the right-hand rotary 

knob is then either assigned to set value of current or power

3.4.6  Manual adjustment of set values

Adjusting set values of voltage, current and power is a key feature of a power supply and hence the two rotary 

knobs on the front of the device are usually assigned to two of the three values, in manual operation. Default as-

signment is voltage and current.
During manual operation, these set values can only be adjusted with the 

rotary knobs

.

Adjusting a value is immediately committed, no matter if the output is switched on or off.

When adjusting the set values, upper or lower limits may come into effect. See section “3.4.4. 

Adjustment limits”. Once a limit is reached, the display will show a note like “Limit: U-max” etc. 

for ~2 seconds.

 

How to adjust values U, I or P with the rotary knobs

1. 

First check whether the value to be changed is assigned to one of the rotary knobs already. The assignment 

can be changed by selecting a different view mode. See 

„3.4.5. Display modes for actual and set values“

.

2. 

With mode 

UI

 selected and as long as the main display is active, turn the left-hand knob to adjust output 

voltage and the right-hand knob to adjust the output current. In mode 

UP

, turn the right-hand knob to adjust 

the output power, as long as the power set value with unit W (Watts) is displayed. The arrow buttons 

 

 can be used to toggle between current and power set value.

3. 

Any set values can be adjusted with the adjustment limits. For switching the digit to adjust, push the rotary 

knob that you are currently using to adjust the value. Every push moves the cursor under the digit in clock-

wise order: 

Содержание PS 9080-100 1U

Страница 1: ...THE POWER TEST EXPERTS User Manual Attention This document is only valid for devices with firmware KE 2 07 and HMI 2 03 or higher For availability of updates for your device check our website or cont...

Страница 2: ......

Страница 3: ...ting the USB port 26 2 3 11 Initial commission 26 2 3 12 Initial network setup 27 2 3 13 Commission after a firmware update or a long period of non use 27 3 OPERATIONANDAPPLICATION 3 1 Important notes...

Страница 4: ...tepro Systems LP Specifications subject to change without notice 4 3 Calibration readjustment 51 4 3 1 Preface 51 4 3 1 1 Preparation 51 4 3 2 Calibration procedure 51 5 CONTACTAND SUPPORT 5 1 General...

Страница 5: ...document 1 1 4 Explanation of symbols Warning and safety notices as well as general notices in this document are shown in a box with a symbol as follows Symbol for a life threatening danger Symbol for...

Страница 6: ...of the device in Volt Series 8 Series 8000 or 800 9 Series 9000 Type identification PS Power Supply usually programmable PSI Power Supply Intelligent always programmable ELR Electronic Load with Reco...

Страница 7: ...equipment and to the source ESD regulations must be applied when plugging interface cards or modules into the relative slot Interface cards or modules may only be attached or removed after the device...

Страница 8: ...cleaning of the equipment ensure that all employees who use the equipment have read and understood the manual Furthermore the users are to be regularly schooled in working with the equipment and the...

Страница 9: ...hold was lower than the actual output voltage Critical The device and or the load could be damaged Signal OCP OverCurrent Shutdown of the DC output due to excess of the preset limit Non critical prote...

Страница 10: ...1 7 ms 1 4 ms 2 2 ms 2 ms 2 ms Display Resolution See section 1 9 4 4 Resolution of the displayed values Display Accuracy 3 0 2 UNom 0 2 UNom 0 2 UNom 0 2 UNom 0 2 UNom Ripple 2 100 mVPP 5 2 mVRMS 29...

Страница 11: ...1150 V DC Miscellaneous Cooling Temperature controlled fans front inlet rear exhaust Ambient temperature 0 50 C Storage temperature 20 70 C Humidity 80 not condensing Standards EN 61010 EN 61326 Overv...

Страница 12: ...s 2 2 ms 2 ms 2 ms Display Resolution See section 1 9 4 4 Resolution of the displayed values Display Accuracy 3 0 2 UNom 0 2 UNom 0 2 UNom Ripple 2 76 mVPP 4 2 mVRMS 234 mVPP 40 mVRMS 156 mVPP 26 mVRM...

Страница 13: ...V DC 1150 V DC Miscellaneous Cooling Temperature controlled fans front inlet rear exhaust Ambient temperature 0 50 C Storage temperature 20 70 C Humidity 80 not condensing Standards EN 61010 EN 61326...

Страница 14: ...cifications subject to change without notice Figure 1 Front side Figure 2 Rear side 1 8 4 Views A Mains switch D Exhausts G AC input connector B Control panel E Share Bus and remote sensing connectors...

Страница 15: ...15 PS 9000 1U Series PS 9000 1U Series DC Laboratory Power Supply User Manual Figure 3 Left hand side view with DC cover Figure 4 Right hand side view without DC cover...

Страница 16: ...16 PS 9000 1U Series 2015 Intepro Systems LP Specifications subject to change without notice Figure 5 Top view...

Страница 17: ...ubmenus and parameters direction up left or changes view mode Button Navigates through menus submenus and parameters direction down right or changes view mode Button Submits altered parameters or set...

Страница 18: ...to 1500 V DC In addition the devices offer as standard the possibility for parallel connection in Share bus operation for constant current sharing Operating in this way allows up to 10 units to be com...

Страница 19: ...pushing the corresponding rotary knob in order to move the cursor The left hand rotary knob is always assigned to the DC output voltage and related parameters like OVP or U max whereas the right hand...

Страница 20: ...ts The number of decimal places depends on the device model The values have 4 or 5 digits while actual and set values related to the same physical unit always have the same number of digits Adjustment...

Страница 21: ...other reserved ports The standard port for this device is 5025 default IP is 192 168 0 2 Via TCP IP and this port communication to the device can be established in most of the common programming lang...

Страница 22: ...ore commissioning the equipment should be visually inspected for damage and completeness using the delivery note and or parts list see section 1 9 3 Scope of delivery An obviously damaged device e g l...

Страница 23: ...et or multi socket outlet as long as the socket features a safety contact PE and is capable for 16 A When connecting the device to a multi socket outlet along with other electric devices it is importa...

Страница 24: ...erage ambient temperature up to 50 C we recommend up to 10 A 0 75 mm AWG18 up to 15 A 1 5 mm AWG14 up to 30 A 4 mm AWG10 up to 40 A 6 mm AWG8 up to 60 A 16 mm AWG4 up to 100 A 25 mm AWG2 per lead mult...

Страница 25: ...ad The device recognizes the remote sensing mode automati cally and regulates the output voltage only in CV operation at the load rather than at its own DC output In the technical specifications see s...

Страница 26: ...it is strongly recommended to use and install the included driver installer on the USB stick to gain maximum compatibility of the device to our software 2 3 10 2 Driver installation Linux MacOS We ca...

Страница 27: ...P http 192 168 0 2 or the default host name http Client only possible if there is a running DNS in the network into the URL box 3 After the website has been completely loaded check the status field it...

Страница 28: ...output power according to P UOUT IOUT reaches the adjusted current or power limit In both cases the device will automatically change to constant current or constant power operation whichever occurs fi...

Страница 29: ...current according to I P U The adjustable set value of current as indicated in the display is always a upper limit While the DC output is switched on and constant power mode is active then the conditi...

Страница 30: ...witch on again to continue operation 3 3 3 Overvoltage An overvoltage alarm OVP will switch off the DC output and can occur if the power supply itself as a voltage source generates an output voltage h...

Страница 31: ...red 3 4 2 Switching off the device On switch off the last output condition and the most recent set values are saved Furthermore a PF alarm power failure will be reported but can be ignored The DC outp...

Страница 32: ...min ____A I max ____A P max ____W General Calibration Reset Communication IP Settings 2 HMI Setup Language English Deutsch Display Language Brightness 1 10 HMI Lock HMI lock Lock all ON OFF possible E...

Страница 33: ...o switch off DC output after PF alarm 6 Determines how the DC output shall react after a power fail PF alarm has occurred OFF DC output will be switched off and remain until user action AUTO DC output...

Страница 34: ...IP address in standard IP format Subnet mask Only available with setting Get IP address Manual Default value 255 255 255 0 Permanent manual setting of the subnet mask in standard IP format Gateway Onl...

Страница 35: ...e That also includes remote control via analog or digital interface In remote control the global range of 0 100 digital resp 0 5 V 0 10 V remains only narrowed by the limits defined here An example yo...

Страница 36: ...after leaving the setup menu and the right hand rotary knob is then either assigned to set value of current or power 3 4 6 Manual adjustment of set values Adjusting set values of voltage current and p...

Страница 37: ...tput on during manual operation or digital remote control can be disabled by pin REM SB of the built in analog interface For more information refer to 3 4 3 1 and example a in 3 5 4 6 How to manually...

Страница 38: ...via the digital interface is active Remote then remote control is immediately terminated and must be reactivated at the PC once Local is no longer active If remote control is via the analog interface...

Страница 39: ...value outputs CMON VMON Input of excess set values e g 5 V in selected 5 V range or 10 V in the 10 V range are clipped by setting the set value at 100 Before you begin please read Important notes for...

Страница 40: ...5 V correspond to 0 100 von PNom Accuracy 0 5 V range 0 4 Accuracy 0 10 V range 0 2 Input im pedance Ri 40 k 100 k 9 VMON AO Actual voltage 0 10 V or 0 5 V correspond to 0 100 von UNom Accuracy 0 2 at...

Страница 41: ...n 2 03 this is also available without remote control being active It is recommended that a low resistance contact such as a switch relay or transistor is used to switch the pin to ground DGND The foll...

Страница 42: ...ated Pin Remote LOW The set values VSEL and CSEL are generated from the reference voltage VREF using potentiometers for each Hence the power supply can selectively work in current limiting or power li...

Страница 43: ...in REM SB to the level that corresponds to DC output off then switch it on again See section 3 5 4 6 Application examples for levels and logic How to acknowledge an alarm in the alarm buffer status du...

Страница 44: ...between a simple lock and a PIN lock which requires to enter the correct PIN every time someone wants to unlock the HMI How to lock the HMI 1 Switch off the DC output and push button to call the setup...

Страница 45: ...nge in HMI language On calling up the menu page and selecting a profile the most important settings can be seen but not changed How to save the current values and settings work profile as a user profi...

Страница 46: ...cted to the next unit using cables with cross section according to the maximum current and with short as possible length 3 9 1 2 Wiring the Share bus The Share bus is wired from unit to unit with an i...

Страница 47: ...same kind and model i e power supply with power supply like for example PS 9080 100 1U with PS 9080 100 1U or similar models like PS PSI 9080 120 2U Series connection is not supported by the software...

Страница 48: ...ed limits for the application should be set in the power supply This can be done via the interface The electronic load should preferably be in operating mode CV The load will then using the Share Bus...

Страница 49: ...put s of all units switched off In this combination of devices it is recommended to always switch on the DC output of the source first and then the DC input of the sink 1 Discharge of the battery to 2...

Страница 50: ...ipping destination is abroad the necessary customs papers are attached 4 2 1 Replacing a defect mains fuse The device is protected by one fuse T16 A 250 V 35 mm which is located inside the device This...

Страница 51: ...imeter for voltage with a max error of half the PS s voltage error That measure ment device can also be used to measure the shunt voltage when readjusting the current If the current is also going to b...

Страница 52: ...and start to measure the output current I mon Simply confirm here with 4 After this the device requests you to enter the output current which you are measuring with the shunt at Measured value Enter i...

Страница 53: ...4 3 2 2 Actual values Actual values of output voltage with and without remote sensing and output current are calibrated almost the same way as the set values but here you don t need to enter anything...

Страница 54: ...86 sales inteproate com www inteproate com United Kingdom Intepro UK Ltd 9 Lakeside Business Park Swan Lane Sandhurst Berkshire GU47 9DN UK Tel 44 012 5287 5600 China Intepro Power Block 7 Fourth Indu...

Отзывы: