EN – ENGLISH
7
4
Reinstall the previously removed screw through the hole in the retainer and into the
screw hole located directly beneath the AC power entry module.
Once the retainer
is installed, the AC power cord cannot be removed from the ICP EXPRESS Monitor
without first removing the retainer
ICP EXPRESS Cable
The ICP EXPRESS Monitor is supplied with one ICP EXPRESS Cable for connecting to
Codman Microsensor ICP Transducers. See Figure 8.
CAUTION: Use only the Codman
ICP EXPRESS Cable to connect the Codman Microsensor to the ICP EXPRESS
Monitor
The ICP EXPRESS Cable is also available separately (Catalog No. 82-6636).
The ICP EXPRESS Cable is reusable and is supplied nonsterile. Refer to the Sterilization
section for details.
CAUTION: DO NOT AUTOCLAVE the ICP EXPRESS Cable Sterilize
the ICP EXPRESS Cable using ethylene oxide (EtO) gas only
The ICP EXPRESS Cable has a non-volatile memory element (EEPROM) embedded
in-line which is used to store the transducer zero reference value of any attached
Codman Microsensor. The ICP EXPRESS Monitor automatically stores the zero
reference value in the cable when the transducer zeroing process is complete. By
storing the transducer zero reference value in the cable, the user will not have to
manually adjust the offset if the patient is moved to another monitor or to another
ICP EXPRESS unit, as long as the original ICP EXPRESS Cable is retained with the
equipment in use. If the original ICP EXPRESS Cable is not retained when switching to
new equipment, the user will be required to manually adjust the stored zero reference
value via the manual zero referencing function of the ICP EXPRESS Monitor.
The ICP EXPRESS Monitor will erase the previously stored zero reference value and
store a new zero reference value in the ICP EXPRESS Cable whenever a new Codman
Microsensor is zeroed, or when the user chooses to adjust the reference manually.
The memory element in the cable is designed to perform over one million erase/write
cycles and will retain data for several years.
NOTE:
A transducer zero reference value
of “-99”, as displayed by the ICP EXPRESS Monitor during a manual zeroing or zero
reference verification procedure, is an indication of an ICP EXPRESS Monitor EEPROM
read failure. This may be a result of a simple random timing error by the ICP EXPRESS
Monitor and can be corrected by selecting the ADJUST REFERENCE menu item and
manually adjusting the displayed zero reference value to the correct zero reference
value. However, this error may also be an indication of a failure in the ICP EXPRESS
Cable EEPROM circuit due to deterioration associated with age or mechanical abuse.
CAUTION: Discontinue use of any ICP EXPRESS Cable that continues to yield a “-99”
transducer zero reference value (as displayed on the ICP EXPRESS Monitor screen)
during the manual zeroing or zero reference verification procedures
CAUTION: DO NOT immerse the ICP EXPRESS Cable in any liquid, or allow liquid to
enter the plug or connectors Do not use cable if connectors become wet Erratic
results can occur
CAUTION: Verify cable integrity prior to use Do not use any cable that fails to meet
the pin-to-pin resistance or electrical leakage specifications
The ICP EXPRESS Cable integrity must be evaluated by qualified technical personnel.
The cable must be routinely inspected for pin-to-pin resistance and electrical leakage.
Pin-to-pin resistance
can be measured using an ohmmeter. Resistances between pins
1–1, 2–2, 3–3, and 4–4 must be one ohm or less. See Figure 8.
Electrical leakage
can be evaluated using an ohmmeter. Measure the open circuit
resistance between pins 1, 2, 3, 4 (inclusive) and all other pins accessible on the ICP
EXPRESS Monitor connector end. The open circuit resistances must be 300 MΩ or
greater.
Technical Specifications
Pressure Channel
Operating Range:
–50 to 250 mmHg
Transducer Excitation Voltage:
5 VDC, ±2.6%
Transducer Sensitivity (Gain):
5
µ
V/V/mmHg
Input Impedance:
10 MΩ
Frequency Response:
0 to 200 Hz (–3 dB)
Zero Range:
±100 mmHg (auto pushbutton
Drift:
<0.1 mmHg/°C (exclusive of ransducer)
Pressure Display Accuracy:
±1% of reading or 1 mmHg, whichever is
larger (exclusive of transducer)
Pressure Display Resolution:
1 mmHg
Isolation:
Isolated from ground related circuits
by >4000 Vrms.
Leakage Current:
<10
µ
A at 120/127 VAC, 60Hz
Alarms
Alarm Parameter:
Mean ICP
Low (LO) Limit, Default:
0 mmHg (user adjustable)
High (HI) Limit, Default:
20 mmHg (user adjustable)
Low (LO) Limit Range:
–50 to +249 mmHg; 1 mmHg steps
High (HI) Limit Range:
–49 to +250 mmHg; 1 mmHg steps
External Patient Monitor Interface
Output Signal Sensitivity:
5
µ
V/Vex/mmHg or 40
µ
V/Vex/mmHg (user
selected; referenced to excitation voltage
of patient monitor)
Output Signal Accuracy:
±1 mmHg or 1%, whichever is greater
(exclusive of transducer)
Reference Signals:
0, 20 and 100 mmHg
Input Impedance:
350–500 Ω
Source Impedance:
350 Ω
Frequency Response:
0 to 200 Hz (–3 dB)
Recommended External Patient Monitor Specifications
Excitation Voltage Range:
0.5 to 10.0 VDC or VAC rms
Excitation Voltage Frequency:
DC to 5000 H
Transducer Sensitivity (Gain):
5
µ
V/Vex/mmHg or 40
µ
V/Vex/mmHg
Drift:
<0.1 mmHg/°C
Zero Range:
at least ±100 mmHg
Frequency Response:
at least 10 Hz
Leakage:
<10
µ
A at 120 VAC, 60 Hz
Isolation:
Isolated from ground related circuits by
4000 Vrms.
Display
Type:
Liquid Crystal Display (LCD)
Active Viewing Area:
2.215 in. x 1.51 in. (5.6 x 3.8 cm)
Matrix Size:
128 x 64 pixels
Backlight:
Blue-Green Electroluminescent (EL)
Viewing Angle:
±30°
Environmental
Operating Temperature Range:
5°C to 45°C
Non-Operating Temperature Range
(Storage and Transportation):
–40°C to 60°C
Operating/Non-Operating
Humidity Range:
5% to 90% (non-condensing)
Operating Atmospheric Pressure
Range:
700 hPA to 1060 hPA
Содержание Codman ICP Express
Страница 2: ...This page is intentionally left blank...
Страница 6: ...This page is intentionally left blank...
Страница 72: ...67 This page is intentionally left blank...
Страница 73: ...68 This page is intentionally left blank...
Страница 74: ...69 This page is intentionally left blank...
Страница 75: ...70 This page is intentionally left blank...