34
INTAB Interface-Teknik AB
3.2.1 Opto Isolated
The opto-isolated inputs can be used in two ways. Which method you use
depends on your particular application.
See figure!
The transducer marked ”PULSE” has to be able to drive the input:
see 4.2.1 for levels and frequencies.
Remember that mechanical
contacts
always
bounce
when brought together.
This bouncing will, unless filtered, result in a burst rather than one pulse.
VO
GND
OR
DI b
DI a
PULSE
MAX 24V
OPTO ISOLATION
ANY POLARITY
DI a
DI b
3.2.2 TTL Compatible
The TTL-compatible inputs (P = Pulse in) are connected as in the figure.
Note that signal ground (GND) of the digital card has to be connected to
signal ground of the system or transducer that is used with these inputs.
1
2
TTL/CMOS
AND/OR
GND
GND
P2
P1
Vp
Vp