24
INTAB Interface-Teknik AB
8 TECHNICAL
SPECIFICATIONS
8.1 Analog
Data
8.1.1 Method
Σ
-
∆
converter
8.1.2 Measurement interval
Programmable in 1s steps - min 1s
N.B. : All channels are measured within the same second
8.1.3 Time Base
X-tal reference
Uncertainty
20ppm at 20°C
Stability
50ppm (-20°C to +60°C)
8.1.4 Inputs
8,16 or 24
differential
inputs.
Impedance:
Overvoltage protection:
Min 400 k
Ω
.
30V continuous.
Higher voltages may damage the inputs
N.B.: current inputs are shunted by 51
Ω
8.1.5 Ranges
Preset:
0-20mA, +/-10V, +/-1000mV, +/-100mV
or
+/-50 mV
Temperatures: 50mV thermo emf
Thermoelectric voltages are linearized and cold junction
compensated