XI.

DIAGNOSTIC TESTS AND 2ND FUNCTION KEY

To access the Diagnostics and 2nd Functions, the unit must be in the “Ready” mode,

press 

2ND FUNC, 

then enter the number of the diagnostic test desired. To exit the

diagnostic test, press 

2ND FUNC a

gain.

Diag. 0 - Custom User Label Programming -

 THIS IS NOT A TEST!

Refer to Section VIII, 

“Programming the Display, ” 

for information.

Diag. 1 - Keypad Test -

 In this test, pressing each key on the front panel keypad will

result in the alphanumeric display showing the name of the key pressed. The numeric
display will show 0000 for keys that are not numbers, and will display the numeral (for
example: 1111, 2222, etc.) when the number keys are pressed. If any key does not
produce the proper display, that key may not be functioning. Contact the factory if this
occurs.

Diag. 2 - Display Test -

 This test is to make sure that all of the LED segments on the

displays and all of the LED indicator lights are working correctly. During the test, the
Numeric display will scroll through all the numbers (i.e. 

1111,

 2222, 3333, etc.) an d

will place the decimal point in each possible location. The alphanumeric display will
also display all of the numbers (i.e. 

11111111,

 22222222, etc.). The six LEDs on the

front of the unit indicating switch position and operating condition will light up one at a
time. If a segment is missing, or an LED indicator light fails, contact the factory.

Diag. 3 - Input Switch Test -This 

is a test for each of the remotely wired inputs.

Each possible input 

(Local/Remote, Keypad Lockout, Forward/Reverse, etc.)

is represented by a 0 or a 1 on the alphanumeric display. The 

E-Stop, Run, Jog, 

and

Stop inputs are represented by a 0 or a 1 on the speed display. Toggle the input to
check whether or not it is properly wired. When the digit corresponding to that input is
displayed as 0 the switch is Open, when it is displayed as 1 it is activated. E-Stop and

Stop 

are 1 when opened and 0 when closed. 

Left to right,

 the eight digits on the alpha-

numeric display (00000000) correspond to the following inputs: Closed/Open Loop;
Master/Follower; Forward/Reverse; Keypad Unlocked/Locked out; Set Point 1(3)/2(4);
Set Point 3 & 4 Disabled/Set Point 3 & 4 Enabled; No Scroll/Scroll Down; No Scroll/
Scroll Up. Left to right, the four digits on the numeric display (0000) correspond to the

following command inputs: E-Stop; Run; Jog; Stop.

Diag. 4 - Memory Test -

 This is an internal test by the micro-processor on its

memory locations. The alphanumeric display will show “Test ROM” for a few seconds,
then it will show either, “ROM ok” or “Fail.” If the test fails, contact the factory.

Diag. 5 - Alarm Output Test -

 This is a test of the Status and Alarm outputs. In this

test, pressing a 1 through 6 on the keypad will result in completion of one of the Output
circuits. When in this diagnostic, the alphanumeric display will read “Output” until a
key (1 through 6) is pressed, then it will name the output being tested and the circuit
will be completed. A light wired into the circuit will illuminate, a relay will actuate, or
an alarm will sound, depending on how the circuit is wired. After an output is tested,

press the STOP button to turn the output off. To exit the Diagnostic, press the Stop, then

23

the 2ND FUNC key. The following numbers correspond to the following outputs:

1 = Ramp Complete; 2 = Zero Speed; 3 = High Alarm; 4 = Low Alarm; 5 = Deviation

Alarm; 6 = Drive Enable

Diag. 6 - Communications Test -

 This test is used to test the RS422 hardware. To

perform the test, place a jumper wire between the Send and Receive lines at the terminal
(RXD- to TXD- and RXD+ to TXD+). The MicroSpeed 196 tests its communication
circuitry to make sure it is functioning. If it is, the alphanumeric display will read “Pass”
and the numeric display will name the node address. If it is not functioning properly, the
alphanumeric display will display “Fail.”

Diag. 7 - Reset to Factory Programming -

This 

is Not a Test. Pressing 

2nd

FUNC,

 

then accessing Diag. 7 will reset the programming in memory to the factory

defaults (see Appendix I for factory default settings). The alphanumeric display will read
“Reset”.

Diag. 8 -

 Not Used.

Diag. 9 - Digital Speed Pot -

 This function is explained in Section IX 

“Calibrating

the Isolated Command Output to the Drive”

.

It can be used as a test of the analog output.

24

Содержание MicroSpeed 196

Страница 1: ...MSMAN32C MicroSpeed 196...

Страница 2: ...is designed in sequential order from basics to the highest level of MicroSpeed 196 programming The deeper you read into each section the less relevent the options may be to your application If this is...

Страница 3: ...d to the wiper of the speed potentiometer The isolated common terminal on the MicroSpeed 196 is connected to reference common on the drive Terminal locations are as follows TB2 2 Isolated Command Outp...

Страница 4: ...ignal from a PLC The MicroSpeed 196 only reads this input while stopped in the Ready mode Wiring of the Master Follower input is as follows TB1 15 Master Follower to common Open Master Closed Follower...

Страница 5: ...puts relays LEDs etc Power to the output circuits should be wired parallel with the Diode Protect TB2 20 terminal if inductive loads are driven to avoid damage to the output circuitry See Diagram 5 Wi...

Страница 6: ...display will prompt for a variable number and the numeric display will indicate two zeros 00 which will reflect the variable number entered Enter the desired variable number and press ENTER If you are...

Страница 7: ...FOLLOWER MODE The Microspeed 196 in the Follower mode mirrors a lead frequency input and can be scaled by the Ratio Set Point The lead frequency is typically generated from a digital transducer on ano...

Страница 8: ...can be independently locked The section of the keypad that is locked is determined by placing a 1 one in the appropriate location Any section or combination of sections can be locked See chart below...

Страница 9: ...ART MIN 56 PARTSIHR 57 Percent 58 RATIO 65 Customized 15 Diagnostic 0 Custom User Unit Labels If the choices presented in Variable 10 do not provide the display that is needed a Custom User Unit Label...

Страница 10: ...ect the proper polarity of the Command Output Voltage using the Forward Reverse Input Open Positive Closed Negative 3 Access Diagnostic 9 Digital Potentiometer by pressing 2nd Function key then the 9...

Страница 11: ...ters at the unstable speed 19 FOLLOWER MODE PID CONSIDERATIONS Variable 09 Max Lead Wind Up If you are speed matching and have very fast master speed changes or if you are bringing the Follower motor...

Страница 12: ...ill decrease motor speed Variable 28 Auxiliary Input Delay Delays the start of the Auxiliary Trim function on startup Enter the number of 10 msec update cycles to wait When ENTER is pressed the MicroS...

Страница 13: ...g inputs Closed Open Loop Master Follower Forward Reverse Keypad Unlocked Locked out Set Point 1 3 2 4 Set Point 3 4 Disabled Set Point 3 4 Enabled No Scroll Scroll Down No Scroll Scroll Up Left to ri...

Страница 14: ...oSpeed 196 with node address 01 will be the only MicroSpeed 196 to answer the host computer but all MicroSpeed 196 s will respond to global commands Never duplicate an address on the same communicatio...

Страница 15: ...ommunication Valid characters from the host are 0 command 1 read variable or 2 write variable The MicroSpeed 196 will echo the host in this character location unless the host communication message has...

Страница 16: ...in a message it will respond with a 3 in the Character 4 position The type of error will be specified in the Character 6 position Error type Problem 1 Parity Error Standard Definition 2 Data Error Da...

Страница 17: ...aximum Speed PPR Maximum Lead RPM Lead PPR User Unity Ratio Master Follower The active set point will only change with a Change Set Point command A write command to the active set point while running...

Страница 18: ...mp Compete Zero Speed Deviation Drive Enable Operational Specifications Accuracy Respons Modes of Operation Preset Speeds Accel Decel Times Processor Speed Display Alphanumeric Display Function LEDs D...

Страница 19: ......

Отзывы: