Heel and Foot Sensors

There are two types of load bearing sensors:

(1) 

The clinician heel sensor is used during data collection and analysis while testing a 

potential WalkAide user.

(2) 

The optional full-length foot sensor is sent home with users whose gait pattern does 

not provide sufficient tilt information to reliably trigger the stimulation with the tilt sensor.

Modified Parameters

Walking parameters that are associated with or derived from the WalkAide unit, Heel Sensor 

and/or Hand Switch after data collection and processing have occurred (e.g. Zoom, Autoset, 
Optimize, or manual adjustments).

Create Patient Profile

Creates a new file for a new patient.

Open a Patient Profile

Opens an existing file for an existing patient.

Optimize Gait Program

Function of the WalkAnalyst software that computes features of the collected walking data 

and configures the ON and OFF thresholds and other parameters to optimize the timing and 

duration of the stimulation for the individual’s gait.

Peripheral Nerve Stimulator

Determines the viability of the nerve to muscle pathway and allows accurate location of the 

peripheral nerves, specifically the superficial and deep branches of the common peroneal 

nerve, for placement of electrodes.

Adjustment

Settings that allow Rapid Follow-Up-Fine adjustments.

Rapid+ Programming

Rapid data collection and programming.  Default settings are Tilt mode.

Reset Usage Log

Resets the “Hours Per Day” and “Stims Per Day” tracking functions.

Reset Zoom/Unselect

Reverses selection function, returning to the original data set and entire walking sequence.

Retrieve Usage Log

Downloads Usage Log following fitting/programming of the WalkAide - First 69 days of “Hours 

Per Day” and “Stims Per Day” information since log was last cleared.

Restore Original

Reverses Autoset / Optimize functions and manual adjustments, returning thresholds and 
control times to original values.

 

Save and Analyze

Function that saves and adds the current walking trial as a date and time stamped event 
under Collected Walking Trials.

Save Copy

Saved WalkAide settings listed as date and time stamped events.

Send to WalkAide

On the walking data graph screen or on the modified parameter screen, allows the clinician 

to send parameters to the WalkAide unit using the WalkAnalyst software and Bluetooth 
connection.

STIM

Abbreviated form of the word stimulation.

Stimulation Mode

Allows the clinician to select from Heel, Hand or Tilt to trigger the stimulation during walking.

Stimulus Setting

Allows the clinician to adjust the characteristics of the pulses within the stimulus train.  
Adjustments include Pulse Width, Time Between and Extra Stimuli functions.

Usage Logs

Collects WalkAide usage information, specifically Hours Per Day and Stims Per Day, for 69 

days of walking activity since the usage log was last cleared.

Walking Data Thresholds

Determine the timing of the initiation and termination of the stimulation for either the Heel or 
Tilt sensors.

WalkAide

A battery-operated, single channel electrical stimulator that can be used for both therapeutic 
and functional electrical stimulation.

WalkAide Diagnostic Codes

Six digit readout indicating WalkAide System hardware’s internal status / fault.

WalkAide Settings

Walking parameters currently programmed into the WalkAide unit.

WalkAide Serial Number

Automatically logs and registers the serial number from the WalkAide unit on the WalkAide 
Parameters screen.  The serial number can also be found in the battery compartment of the 
WalkAide.

WalkAnalyst

Software used by the clinician to interface with the WalkAide unit.  This allows data collection, 

Содержание The WalkAide System

Страница 1: ...physician LM06 R6 2013 Innovative Neurotronics All rights reserved All trademarks and registered trademarks are the property of their respective holders No part of this manual may be reproduced in an...

Страница 2: ...9 4 1 Configuring initial set up process Pairing 19 5 0 Fitting Process 20 5 1 Pre screening with the Peripheral Nerve Stimulator 20 5 2 Electrode Placements and System Preparations 22 5 3 Fitting a N...

Страница 3: ...ions Do not use on persons with implanted demand type cardiac pacemakers or defibrillators Do not place the electrodes in the carotid sinus region throat Laryngeal or pharyngeal spasms may occur when...

Страница 4: ...to operate the equipment 1 5 Precautions Heart Disease Use caution in applying electrical stimulation to persons suspected of having heart disease More clinical data is needed to show that such person...

Страница 5: ...Glossary AutosetWalkAide Parameters This calculation process adjusts the threshold settings to more closely match the individual data collected Audible Beep Optional Biofeedback Feature Audible signal...

Страница 6: ...d Stims Per Day information since log was last cleared Restore Original Reverses Autoset Optimize functions and manual adjustments returning thresholds and control times to original values Save and An...

Страница 7: ...video One free USB port for the Bluetooth Adapter Windows XP with SP2 or Windows 7 A peripheral nerve stimulator can be purchased separately The peripheral nerve stimulator uses a 9 volt battery to op...

Страница 8: ...mode A non audible alarm with only a red blinking light indicates hibernation mode Establish a Bluetooth connection with WalkAnalyst to automatically clear hibernation mode If the hibernation mode do...

Страница 9: ...Washing Instructions To wash the WalkAide Cuff fabric liner first remove the electrodes and then remove the liner from the cuff Do NOT remove the Black and Red Electrode Locators Make sure to Hand Wa...

Страница 10: ...click on Setup exe and choose Run as Administrator Follow the set up instructions that will appear 4 The WalkAnalyst program will be installed in the Program Files Innovative Neurotronics directory un...

Страница 11: ...to the WalkLink This is called the pairing process Following the guidelines below to accomplish this process Plug in your USB Bluetooth adapter with the computer powered ON Start the WalkAnalyst appl...

Страница 12: ...ot close to a neutral alignment 3 Clean the skin in the area around the head of the fibula with soap and water and wipe dry Failure to adequately prepare the skin may cause improper contact and provid...

Страница 13: ...de lead cable Make sure the BLACK lead negative is connected to the BACK electrode and the RED lead positive is connected to the FRONT electrode Figure 26 5 Turn the WalkAide ON by turning the blue In...

Страница 14: ...nt Profile Add basic information patient ID location primary diagnosis etc and save the patient file Select Program During the initial fitting you have the following programming options Rapid Program...

Страница 15: ...rrow Upon releasing the mouse the WalkAide settings are calculated and an optimization error is displayed An optimization error of less than 20 is recommended Press the Next button to program the Walk...

Страница 16: ...e Log to set the internal date time stamp and start a new 69 day tracking sequence For an existing user Select Retrieve Usage Log to download the most recent walking activity data This Usage Log will...

Страница 17: ...the Data Data processing involves following the sequence of icons from left to right displayed above the Tilt data graph Zoom Autoset and Optimize are the first three steps involved in processing the...

Страница 18: ...cise settings Figure 41 Usage Logs Select Usage Logs tab to view and or clear usage data Figure 42 Preferred Setting You can use the checkbox Set as Preferred Setting to tag a setting as preferred set...

Страница 19: ...ual Users should proceed through the daily wearing schedule and increase wearing time only if no skin irritation and or muscle soreness is present Figure 55 ON Time OFF Time Days 1 3 15 60 minutes 30...

Страница 20: ...performance improvement and justify medical necessity of WalkAide Evaluation options Collect Walking Data No Device When collecting walking data with no assisted device attach the WalkAide on a cuff...

Страница 21: ...de is to be stored for an extended period of time and not used remove the battery from the battery compartment When the alkaline batteries become depleted please dispose of properly in accordance with...

Страница 22: ...of the leg The stimulation is optimized if it enters at the motor nerve and exits after traveling in the direction of the muscle Electrode position varies with each patient Conductivity is enhanced b...

Страница 23: ...poorly timed on and off cycles that may occur with abnormal tibial motion during swing usually from ataxia Adjustment to the Min Time also addresses periods of faster walking or faster steps when the...

Страница 24: ...e testing to prevent possible overstimulation of the user Recommended for pediatric or more sensitive patients a low Pulse Width may provide a less forceful stimulation and allow for a softer contract...

Страница 25: ...ur methods to re optimize the user s walking program These are outlined below with clinical descriptors of when each method may be most indicated Option 1 Rapid Adjustment Adjust stimulation parameter...

Страница 26: ...trial listed below Save WalkAide Settings and make any clinical comments relevant to this walking program Click on 4 Verify WalkAide Settings and change the Stimulation Mode to Hand Send Parameters to...

Страница 27: ...nutes of non use Turn the WalkLink off for a few seconds and then turn it on again Click on the Bluetooth icon in the upper right corner of the screen and Search for Previously Connected WalkLinks Che...

Страница 28: ...f another Bluetooth driver software is causing the conflict you may need help from your IT support team Call the Innovative Neurotronics technical support team if you have trouble resolving this probl...

Страница 29: ...Reference Part Numbers 20 1000 WalkAide Patient Kit 21 1000 WalkAide Patient Kit Japan 20 1000 WalkAide Patient Kit Reprocessed 20 0100 WalkAide 21 0100 WalkAide Japan 20 0100R WalkAide Reprocessed 2...

Отзывы: