Innovative Neurotronics The WalkAide System Скачать руководство пользователя страница 27

50

51

 The electrodes are not wet enough.

 

 The cuff is not adequately tightened causing it to shift with motion and thereby 

 

compromising electrode/skin connection.  Ensure proper positioning of cuff and tighten the 

strap.  Avoid tight, slim-fitting pants that can cause a pull on the cuff.

 Check skin under electrodes for irritation or open areas.

 

 Ensure total contact with skin.

 

5.  Red light is flashing with 4 beeps every two seconds. 

 

 

Possibilities:

Indicates a fault has occurred in the control module.  Turn OFF the light blue Intensity 

 

Knob by turning counter-clockwise to 0. Wait 5 seconds and then turn ON again to see if 

the green light is flashing. If the red light remains lit, the WalkAide unit must be returned to 

Innovative Neurotronics for repair or replacement.

 Contact Innovative Neurotronics at 888-88ININC (888-884-6462) for shipping instructions.

 

6.  Red light is flashing with an audible beep every minute. 

 

 

Possibilities:

Low battery condition. Change the battery.

 

7.  Red light flashing with audible 2 beeps every two seconds.

 

 

 

Possibilities:

If the unit beeps two times every two seconds, you may be in heel mode.  When in heel 

 

mode, the Heel or Foot Sensor must be connected to the unit.  Check device parameters 
and make changes to ‘Stimulation Mode’ if necessary.

8.  Red light is flashing with no audible alarm.

 

 

Possibilities:

Time clock in the WalkAide unit for usage log has stopped working.  Need to reconnect to 

 

the WalkAnalyst program to Clear WalkAide Log. 

 Internal time clock battery needs replacing. Contact Innovative Neurotronics. 

 

9.  The green light does not flash when the WalkAide is turned ON.

 

 

Possibilities:

Make sure the light blue Intensity Knob is turned to 1 or higher.  There will be a clicking 

 

noise when the knob is turned from 0 to ON.

 A new battery may be needed.

 

9.2  WalkLink Troubleshooting

1.  The left-most LED light does not flash a green light every two seconds

The flashing green light is a power indicator.  If it is not flashing, the device may have 

 

simply timed out after approximately 10 minutes of non-use.  Turn the WalkLink off for a 
few seconds and then turn it on again. Click on the Bluetooth icon in the upper right corner 
of the screen and ‘Search for Previously Connected WalkLinks’.  

 Check and make sure the four AA alkaline 1.5V batteries in the WalkLink are new.  Do not 

 

use rechargeable batteries in the WalkLink.

2.  The WalkLink device did not appear during the WalkAnalyst Bluetooth configuration 

process

Turn the WalkLink off for ten seconds, turn it back on and press search again. 

 

 Confirm that the laptop is using Windows XP Service Pack 2 or better, Windows 7 or Vista.

 

 The WalkLink device may need to be unpaired and re-paired with the laptop.  Follow these 

 

steps:

 Turn on the WalkLink device and confirm the flashing green power light.

 

 Use a ballpoint pen or something similar to press and hold the small 

 

recessed unpair button located above the on/off switch.  Depress and hold 
for one second.

 Turn off the WalkLink, wait 10 seconds, turn on the WalkLink and try the 

 

search process again.

Make sure that the WalkLink is on and search for the WalkLink at least five times.  

 

Sometimes the WalkLink is recognized on later searches.

 If you are having difficulty pairing the WalkLink using the Bluetooth Wizard, please refer 

 

to the Alternate Bluetooth Pairing section of the WalkAnalyst Installation & Bluetooth 

Configuration guide provided with the WalkAnalyst Software; it is also located in the help 

portal of the WalkAnalyst software.

3.  A connection has been established with the WalkLink device during installation 

but now messages from WalkAnalyst appear that WalkAnalyst could not find a 
configured WalkLink.

Turn the WalkLink off and back on, and either choose the first option in the Bluetooth 

 

configuration wizard (i.e. Search for Previously Connected WalkLinks) or simply exit and 

restart the WalkAnalyst software application while WalkLink is still ON.

 Make sure the green light on the WalkLink is blinking every two seconds. The WalkLink will 

 

timeout after approximately 10 minutes if the WalkAnalyst software is not running, even if 
the blue light is blinking on the WalkLink initially.  Turn the WalkLink off for a few seconds, 
then turn it back on and restart the WalkAnalyst application.

Содержание The WalkAide System

Страница 1: ...physician LM06 R6 2013 Innovative Neurotronics All rights reserved All trademarks and registered trademarks are the property of their respective holders No part of this manual may be reproduced in an...

Страница 2: ...9 4 1 Configuring initial set up process Pairing 19 5 0 Fitting Process 20 5 1 Pre screening with the Peripheral Nerve Stimulator 20 5 2 Electrode Placements and System Preparations 22 5 3 Fitting a N...

Страница 3: ...ions Do not use on persons with implanted demand type cardiac pacemakers or defibrillators Do not place the electrodes in the carotid sinus region throat Laryngeal or pharyngeal spasms may occur when...

Страница 4: ...to operate the equipment 1 5 Precautions Heart Disease Use caution in applying electrical stimulation to persons suspected of having heart disease More clinical data is needed to show that such person...

Страница 5: ...Glossary AutosetWalkAide Parameters This calculation process adjusts the threshold settings to more closely match the individual data collected Audible Beep Optional Biofeedback Feature Audible signal...

Страница 6: ...d Stims Per Day information since log was last cleared Restore Original Reverses Autoset Optimize functions and manual adjustments returning thresholds and control times to original values Save and An...

Страница 7: ...video One free USB port for the Bluetooth Adapter Windows XP with SP2 or Windows 7 A peripheral nerve stimulator can be purchased separately The peripheral nerve stimulator uses a 9 volt battery to op...

Страница 8: ...mode A non audible alarm with only a red blinking light indicates hibernation mode Establish a Bluetooth connection with WalkAnalyst to automatically clear hibernation mode If the hibernation mode do...

Страница 9: ...Washing Instructions To wash the WalkAide Cuff fabric liner first remove the electrodes and then remove the liner from the cuff Do NOT remove the Black and Red Electrode Locators Make sure to Hand Wa...

Страница 10: ...click on Setup exe and choose Run as Administrator Follow the set up instructions that will appear 4 The WalkAnalyst program will be installed in the Program Files Innovative Neurotronics directory un...

Страница 11: ...to the WalkLink This is called the pairing process Following the guidelines below to accomplish this process Plug in your USB Bluetooth adapter with the computer powered ON Start the WalkAnalyst appl...

Страница 12: ...ot close to a neutral alignment 3 Clean the skin in the area around the head of the fibula with soap and water and wipe dry Failure to adequately prepare the skin may cause improper contact and provid...

Страница 13: ...de lead cable Make sure the BLACK lead negative is connected to the BACK electrode and the RED lead positive is connected to the FRONT electrode Figure 26 5 Turn the WalkAide ON by turning the blue In...

Страница 14: ...nt Profile Add basic information patient ID location primary diagnosis etc and save the patient file Select Program During the initial fitting you have the following programming options Rapid Program...

Страница 15: ...rrow Upon releasing the mouse the WalkAide settings are calculated and an optimization error is displayed An optimization error of less than 20 is recommended Press the Next button to program the Walk...

Страница 16: ...e Log to set the internal date time stamp and start a new 69 day tracking sequence For an existing user Select Retrieve Usage Log to download the most recent walking activity data This Usage Log will...

Страница 17: ...the Data Data processing involves following the sequence of icons from left to right displayed above the Tilt data graph Zoom Autoset and Optimize are the first three steps involved in processing the...

Страница 18: ...cise settings Figure 41 Usage Logs Select Usage Logs tab to view and or clear usage data Figure 42 Preferred Setting You can use the checkbox Set as Preferred Setting to tag a setting as preferred set...

Страница 19: ...ual Users should proceed through the daily wearing schedule and increase wearing time only if no skin irritation and or muscle soreness is present Figure 55 ON Time OFF Time Days 1 3 15 60 minutes 30...

Страница 20: ...performance improvement and justify medical necessity of WalkAide Evaluation options Collect Walking Data No Device When collecting walking data with no assisted device attach the WalkAide on a cuff...

Страница 21: ...de is to be stored for an extended period of time and not used remove the battery from the battery compartment When the alkaline batteries become depleted please dispose of properly in accordance with...

Страница 22: ...of the leg The stimulation is optimized if it enters at the motor nerve and exits after traveling in the direction of the muscle Electrode position varies with each patient Conductivity is enhanced b...

Страница 23: ...poorly timed on and off cycles that may occur with abnormal tibial motion during swing usually from ataxia Adjustment to the Min Time also addresses periods of faster walking or faster steps when the...

Страница 24: ...e testing to prevent possible overstimulation of the user Recommended for pediatric or more sensitive patients a low Pulse Width may provide a less forceful stimulation and allow for a softer contract...

Страница 25: ...ur methods to re optimize the user s walking program These are outlined below with clinical descriptors of when each method may be most indicated Option 1 Rapid Adjustment Adjust stimulation parameter...

Страница 26: ...trial listed below Save WalkAide Settings and make any clinical comments relevant to this walking program Click on 4 Verify WalkAide Settings and change the Stimulation Mode to Hand Send Parameters to...

Страница 27: ...nutes of non use Turn the WalkLink off for a few seconds and then turn it on again Click on the Bluetooth icon in the upper right corner of the screen and Search for Previously Connected WalkLinks Che...

Страница 28: ...f another Bluetooth driver software is causing the conflict you may need help from your IT support team Call the Innovative Neurotronics technical support team if you have trouble resolving this probl...

Страница 29: ...Reference Part Numbers 20 1000 WalkAide Patient Kit 21 1000 WalkAide Patient Kit Japan 20 1000 WalkAide Patient Kit Reprocessed 20 0100 WalkAide 21 0100 WalkAide Japan 20 0100R WalkAide Reprocessed 2...

Отзывы: