Chapter 7 - Maintenance Menu
The splicer has a function to perform routine maintenance. This section describes how to use the maintenance menu.
① Select [Maintenance Menu].
② Select a function to perform.
Remove old electrode
Install new electrode
Loosen the screw
7.2 Replace Electrodes
As electrodes are worn out during the splicing process, oxide generated on the tips of electrodes should be regularly
eliminated. It is recommended that the electrodes should be replaced after 4500 arc discharges. When the number of arc
discharges reaches a count of 4500, a message prompting to replace the electrodes is displayed immediately after turning
on the power. Using the electrodes without a replacement will result in greater splice loss and reduced splice strength.
Replacement Procedure
① Select [Replace Electrodes] in [Maintenance Menu].
② Instruction messages will appear on the screen to turn off the power. Then turn off the splicer.
③ Remove the old electrodes.
I) Loosen screw located on electrode cover.
II) Take electrodes out of the electrode covers. (Electrodes are fixed in electrode cover)
7.1 Maintenance
VIEW 5 Pro has a built-in diagnostic test function that allows the user to evaluate several critical variable parameters with
only one simple step. Perform this function in case of splicer operation fault.
Operation Procedure
① Select [Maintenance] in [Maintenance Menu]
② Execute [Maintenance], then the following checks will be made.
No.
Check Item
Description
1
LED calibration
Measure and adjust the brightness of the LED.
2
Dust check
Check the optical path for dust or dirt and judges whether they disturb fiber
observation. If contamination exists, press the return button twice to display the
location.
3
Adjust Position
Automatic fiber adjustment
4
Motor Calibration
Automatically calibrate the speed of 6 motors.
5
Stabilize Electrodes
Accurately measure the position of the electrodes by ARC discharge.
6
Arc Calibration
Automatically calibrate the arc power factor and fiber splicing position.