25
The function is enabled and the operator has an opportunity to modify the EXHAUST DILUTION threshold, by
pushing the EXHAUST DILUTION push button momentarily. The threshold for the EXHAUST DILUTION
function will be displayed in the top right (CO
2
) display [all other displays will be blank]. The operator can press the
“
” and “
” push buttons and the EXHAUST DILUTION threshold will increases or decreases its value. When the
operator has the value he wants to use (we recommend 12%), pressing ENTER will save the EXHAUST DILUTION
threshold value and the EXHAUST DILUTION indicator will come on. The system now has EXHAUST
DILUTION active.
To disable the function, when the feature is on (i.e. the EXHAUST DILUTION indicator is lit), press and hold the
EXHAUST DILUTION push button for more than five (5) seconds. When the feature is disabled the EXHAUST
DILUTION indicator will be turned on.