L E A K C H E C K :
This option will run a leak check of the system when YES is selected. Use ▲ or
▼ to select between yes and no.
To run the test select YES and place the red cap on the end of the probe. When
ready to run the test press MODE.
During the test this screen will be displayed. When the test is complete PASSED,
BAD PUMP OR FAILED will display in the lower two windows. Press MODE
or MODE EXIT to exit the test.
F I E L D C A L I B R A T I O N :
Use ▼ or ▲ to select between yes and no. When you have finished your
selection you can exit the setup mode by pressing MODE EXIT or go to the next
calibration step by pressing MODE.
This window is displayed as the system purges the gas lines. When the purge is
complete the next screen appears:
This window is prompting you that the next step will be to enter your tag values.
If you want to proceed with the calibration setup, press MODE. Otherwise, if you
would like to exit without calibrating, press MODE EXIT.
If you want to reset
the analyzer to the factory default, setting press MODE EXIT. See Factory
Calibration Reset below.
The tag values should be entered just as they are on the calibration bottle. You are not required to calculate HC
using the PEF value. Instead enter the propane value and the analyzer will internally convert it to HC. When
tag values for CO and CO2 are entered they are entered as percentages. For HC and NOX values they are
entered as ppm. Any gas you do not want to calibrate should be left with a value of zero. O2 does not get
calibrated.
Use ▲ and ▼ to change the number that is currently flashing. To move to
another digit or field use the ◄ or ► button. You can move to the next window
by pressing the MODE button. After entering the tag values press the MODE
button until the Go in the lower right window flashes. If you want to proceed
with the calibration press MODE otherwise if you would like to exit without
calibrating press MODE EXIT. If you proceed with the calibration the next
screen will appear with the Go flashing.
Leak Test
no
Leak Test
wait
FIE
LD
CAL Time
YES
Pur
Ge
Stan dby
CAL
ENT
TAGS
GO
CAL
HC
0000 00.00
CO
CO2
00.00 00.00
O2
NOX
0000 Go
CAL
Содержание FGA
Страница 1: ...FGA4000XDS Four and Five Gas Analyzer ...
Страница 2: ...1 ...
Страница 4: ...3 ...
Страница 8: ...7 ...
Страница 18: ...17 ...
Страница 34: ...33 ...
Страница 40: ...39 ...
Страница 46: ...45 ...
Страница 48: ...47 ...
Страница 50: ...49 ...
Страница 52: ...51 ...
Страница 54: ...53 ...
Страница 56: ...55 ...
Страница 58: ......