18
Chapter 2 –
Using the Analyzer
W A R M - U P
P R E - W A R M - U P L O C K O U T P H A S E
The system enters the PRE-WARM-UP LOCKOUT phase when power is first applied by pressing the POWER
ON/OFF switch (main front panel). The system goes into this state for five minutes. During PRE-WARM-UP
LOCKOUT, the system turns on all the display segments and indicators. All the digital displays show 8888, the
decimal points and all the indicators are lit. This is referred to as LAMP TEST. The operator should use this phase to
ensure that all lights and display segments are functional. Inoperative display segments can easily lead to inaccurate
visual readings. The pump runs during this phase to purge any residual gases from the sample hose.
During this
phase the operator should have the probe exposed to ambient air and should not be near any exhaust.
No
other modes of operation can be initiated during this phase.
After the PRE-WARM-UP LOCKOUT phase, five-minute time period expires, the unit goes into the WARM-UP
OPERATE phase. The first function in this phase is an Auto Zero. After the auto zero, the system tests itself. If the
system is stable, it will leave the WARM-UP OPERATE phase and move to STANDBY mode. If the stability test
indicates that the system requires more warm-up time,
the system will enter the WARM-UP OPERATE phase.
In the STAND-BY OPERATE mode of the WARM-UP phase, the system will allow the operator to initiate a
measurement. A measurement is initiated in the normal fashion. The system will make the measurement in this mode,
and may be at full accuracy.
In the STAND-BY mode, the operator will also be able to turn the pump on and off by the use of the PUMP switch.
The pump can be used normally in this mode to purge the gas from the probe or any other purpose.
In the STAND-BY mode the operator will be able to enter AUTOZERO mode.
If the operator does not initiate any of the functions allowed in the WARM-UP OPERATE phase, the system will wait
for five minutes, auto-zero itself, and test again for stability. If the system is stable at the end of that time the system
will move into STANDBY mode. If the system is still not stable, the system will remain in the WARM-UP
OPERATE phase, and begin to wait for another five minutes. If during the WARM-UP OPERATE phase the
operator initiates an allowable operation, the five-minute time period will be restarted when the allowable operation is
complete.
The WARM-UP OPERATE phase allows measurements to be made quickly after power-up and only slightly outside
the specified accuracy limits of a fully warmed up unit.
Содержание FGA
Страница 1: ...FGA4000XDS Four and Five Gas Analyzer ...
Страница 2: ...1 ...
Страница 4: ...3 ...
Страница 8: ...7 ...
Страница 18: ...17 ...
Страница 34: ...33 ...
Страница 40: ...39 ...
Страница 46: ...45 ...
Страница 48: ...47 ...
Страница 50: ...49 ...
Страница 52: ...51 ...
Страница 54: ...53 ...
Страница 56: ...55 ...
Страница 58: ......