6.2.2
Display-range auto/manual
•
>
>
The upper limit of the displayed leak rate range can be set manually or automatically.
These settings apply to the bargraph (=bar underneath the digital figures in the
measurement mode and y-axis in the trend mode).
With the upper limit defined here the lower limit is set to a value based on the number
of decades (
Please refer to Chapter
If linear scale is selected, the lower limit is always zero. The upper limit is only a
default value. You can change this on the measurement screen with the Soft Key 6
and 7 if you have chosen manual display ranging.
6.2.3
Time axis
•
>
>
The length of the time axis in trend mode can be changed in steps of 16 … 960 s.
Softkey 2:
Manual
The upper limit of the displayed range can be set manually.
Softkey 3:
Arrow down
.Decrease the upper limit if
manual
is chosen. The minimum value is
10
-11
mbar l/s.
Softkey 6:
Automatic
The limit of the displayed range will be chosen automatically.
Softkey 7:
Arrow up
Increase the upper limit if
manual
is chosen. The maximum value is
10
+3
mbar l/s.
Softkey 8:
Save the settings and return to the previous menu.
Softkey 3:
Arrow down
Decrease the length of the time axis. The minimum value is 16 seconds.
Softkey 4:
Scroll on / off
In „scroll on“ the display is written continiously in trend mode. In „scroll off“
the display will be overwritten after the adjusted action is finished.
Softkey 5:
?
Help
Softkey 7:
Arrow up
Increase the length of the time axis. The maximum adjustable value is 960
seconds.
Содержание UL5000
Страница 102: ...102 Declaration of Conformity iina74e SIX fm technical handbook 1408 Declaration of Conformity ...
Страница 103: ...Declaration of Conformity 103 iina74e SIX fm technical handbook 1408 ...
Страница 105: ...Declaration of Conformity 105 iina74e SIX fm technical handbook 1408 ...
Страница 108: ...3 iina74e1RL fm technical handbook 1408 ...