6.6.5.3
Leak rate filter
•
>
>
>
The kind of the leak rate filter can be chosen. The default value is I•CAL.
I•CAL stands for Intelligent Calculation Algorithm of leak rates. It makes sure that the
signals are averaged in optimized time intervals, based on the leak rate intensity.
I•CAL also eliminates noise peaks which are not related to leak rate signals and
provides unexpected short response times for low leak rate signals.
The algorithm used provide excellent sensitivity and response time and is there for
the recommended setting.
6.6.5.4
Mains Frequency
•
>
>
>
The mains frequency takes the different pumping speed of the scroll pump into
account. The frequency of the mains power supply can be selected. The default
setting is 50 Hz for 230 V and 60 Hz for 115 V.
6.6.5.5
Service interval exhaust filter
Here you can enter the service interval of the exhaust filter.
Softkey 3:
Fixed
A filter with a fixed time constant will be used.
Softkey 7:
I•CAL
I•CAL makes sure that the averaging time is optimal based on the leak rate
level.
Softkey 3:
50 Hz
The UL5000 will be run at a mains frequency of 50 Hz.
Softkey 6:
60 Hz
The UL5000 will be run at a mains frequency of 60 Hz.
Softkey 3:
Down
Decrease of the service interval steps of within 500 hours.
Softkey 5:
?
Help
Softkey 7:
Up
Increase of the service interval within steps of 500 hours. The limit is 4000
hours.
Содержание UL5000
Страница 102: ...102 Declaration of Conformity iina74e SIX fm technical handbook 1408 Declaration of Conformity ...
Страница 103: ...Declaration of Conformity 103 iina74e SIX fm technical handbook 1408 ...
Страница 105: ...Declaration of Conformity 105 iina74e SIX fm technical handbook 1408 ...
Страница 108: ...3 iina74e1RL fm technical handbook 1408 ...