84
Error And Warning Messages
iina
74
e 0
8
.f
m
te
chn
ica
l ha
nd
boo
k
(1
40
8)
E34
24V voltage at MSV board is
too low!
Signal MVPZN on the MSV board is active. 24 V signal voltage is too low,
U < 18.3 V.
•
Fuse F1 on the MSV board has blown.
•
24 V power supply voltage is missing.
Switch off the UL5000!
The missing voltage will cause the exhaust valve on the scroll pump to
close which in turn can lead to a contamination of the vacuum system.
•
Reference voltage UREF on the MSV board XT7/1 is too high, U > 5 V.
E35
Anode-cathode voltage is too
high!
•
MSV board is faulty.
•
Anode-cathode voltage is higher than 130 V.
E36
Anode-cathode voltage is too
low.
•
MSV board is faulty.
•
Anode-cathode voltage is lower than 30 V.
E37
Suppressor voltage reference
value too high!
Signal MFSZH on MSV board is active. Suppressor signal command
variable is too high.
•
Suppressor voltage has a short circuit.
•
MSV is faulty.
E38
Suppressor potential too
high!
Suppressor potential is higher than 363V.
•
MSV board is faulty.
E39
Suppressor potential is too
low
Suppressor potential is lower than 297V.
•
MSV board is faulty.
E40
The anode potential exceeds
its nominal value by over
10%!
The actual anode potential exceeds its nominal value by 10%. The
nominal value can be displayed in the service menu.
•
MSV is faulty.
E41
The anode potential has
dropped below its nominal
value by over 10%!
The actual anode potential has dropped below its nominal value by 10%.
The nominal value can be displayed in the service menu.
•
Air inrush.
•
MSV is faulty.
E42
Nominal value of the anode
potential is too high!
Signal MFAZH on MSV board is active.
•
Anode voltage has been short circuited.
•
Nominal value of the anode voltage is too high. Anode voltage is limited
to about 1,200 V.
E43
Cathode current is too high!
•
Signal MPKZH on MSV board is active. Cathode current is too high, I
> 3.6 A.
•
MSV is faulty.
E44
Cathode current is too low!
•
Signal MPKZN on MSV board is active. Cathode current is too low, I >
0.2 A.
•
MSV is faulty.
W45 Emission for cathode 1 can
not be switched on!
Signal MSIBE on MSV board is not active. Emission for cathode 1 can not
be switched on. UL5000 switches to cathode 2. Please order a new ion
source.
W46 Emission for cathode 2 can
can not be switched
Signal MSIBE on MSV board is not active. Emission for cathode 2 can not
be switched on. UL5000 switches to cathode 1. Order a new ion source.
No.
Displayed Message
Description and possible solutions
Содержание UL5000
Страница 102: ...102 Declaration of Conformity iina74e SIX fm technical handbook 1408 Declaration of Conformity ...
Страница 103: ...Declaration of Conformity 103 iina74e SIX fm technical handbook 1408 ...
Страница 105: ...Declaration of Conformity 105 iina74e SIX fm technical handbook 1408 ...
Страница 108: ...3 iina74e1RL fm technical handbook 1408 ...