tinb21e1 (2016-06)
41
Notes:
1) OP = overpressure indication: display will read over pressure
2) Display auto-ranges between Torr and mTorr at 1 Torr
Examples:
1) Gas used is nitrogen (N
2
). Display shows pressure mea-
surement of 10 Torr. True pressure of nitrogen is 10 Torr.
2) Gas used is argon (Ar). Display shows pressure measure-
ment of 600 mTorr. True pressure of argon is 1 Torr.
3) Gas used is oxygen (O
2
). Display shows pressure measure-
ment of 486 mTorr. True pressure of oxygen is 500 mTorr.
6.2
Display - mbar
The table below shows the displayed readings at various pres-
sures for selected gases when engineering units selected is in
mbar.
Displayed pressure readings vs. true pressure for selected
gases - engineering units in mbar (see following table):