2
Getting Started
Gas Monitoring and Sensors
User Interface (UI)
Preparing the Tango TX2 for First Use
Gas Monitoring and Sensors
Measuring gas at two-second intervals, the Tango® TX2 is a portable, two-gas diffusion monitor
(instrument) for personal protection. See
Tango TX2 is compatible with the Industrial Scientific sensors listed below and supports
only
a two-sensor
configuration. The instrument does
not
support a single-sensor configuration.
Sensors are customer replaceable. To install new or replacement sensors, see
”.
Table 2.1 Compatible gas sensors
Gas sensor
Display screen
symbol
Part number
Details
Carbon Monoxide (CO)
17160204
Carbon Monoxide with low Hydrogen cross-
sensitivity (CO/H
2
Low)
17160208
Hydrogen Sulfide (H
2
S)
17160207
Nitrogen Dioxide (NO
2
)
17160205
Sulfur Dioxide (SO
2
)
17160206
Data are saved every ten seconds and stored to the instrument data log. The data log has these
characteristics:
•
Capacity for approximately three months
of data for a unit that is on 24 hours a day.
•
Date- and time-stamped entries for
60 alarm events, 30
error events, and up to 250 manual calibration
procedures and bump tests.
•
As new data are logged to memory, the oldest data are overwritten.
•
Downloaded on docking in a compatible docking station from Industrial Scientific.
User Interface (UI)
As illustrated below, Tango TX2 UI comprises instrument buttons, lights, sounds, and vibration along with
display-screen characters and symbols. The home-screen illustration indicates how gas readings generally