20
Figure 4.2 Settings menu
Display screen
Setting and options
“EUr”
for Europe
“CR”
for Czech Republic
“AUS”
for Australia
Save ( ) the setting.
Language
Set ( ) the display screen for one of the available languages.
“
EN
”
for English
“
F
”
for French
Save ( ) the setting.
Always on
Set ( ) the instrument to always be on; otherwise, choose disabled.
Note
: If the instrument has a valid security code, it can be used by the
worker to power off an always-on instrument.
0 = disabled
1 = enabled
Save ( ) the setting.
Shutdown in alarm
Use ( ) this setting to prevent the worker from shutting down the
instrument when it is in alarm; otherwise, allow in-alarm shutdown.
0 = prevent shutdown
1 = allow shutdown
Save ( ) the setting.
Deadband
Use ( ) this setting to display a gas reading of
“0” (zero) when the
detected gas
concentration falls within a sensor’s deadband
.
0 = disabled; always display
actual
gas reading
1 = enabled; display
zero
when the detected gas concentration
is within the deadband
Save ( ) the setting.
Alarms when docked
Use ( ) this setting to disable alarms when the instrument is docked;
otherwise allow alarms to occur when docked.
0 = alarms disabled while docked
1 = alarms can occur while docked
Save ( ) the setting.
Gas alert function
Use ( ) a setting of
“enabled”
to make gas alerts functional. Gas
alerts notify the worker when a detected gas concentration may be
approaching alarm levels.
0 = disabled
1 = enabled
Save ( ) the setting.
Go ( ) to the first setting.