7
Service and Warranty
Service
Warranty
Service
Perform all service tasks on a nonconductive surface in a well-lit area that is known to be nonhazardous.
Wear grounding straps to prevent electrostatic discharge (ESD), which can damage the
instrument’s
electronics.
Use care when working with the adhesive-backed filters and gaskets.
✓
Be careful not to pierce or tear these items.
✓
When using tweezers, apply gentle pressure.
✓
Once the adhesive touches a surface, any attempt to remove or reposition the item may cause it
damage.
Use care when working with gas sensors and the case top's water barriers.
✓
Do not touch the membranes as this can contaminate these items.
✓
Use care not to damage the membranes.
✓
Use care not to separate the gas sensor from its membrane.
Supplies
T10 Torx screwdriver (for case bottom and clip screws)
Needle-nose tweezers (for barrier and filter service tasks)
Tango TX2 illustrated
The diagrams below show disassembled views of Tango® TX2 and its case top assembly.
identifies parts, part numbers, and whether an item is customer-replaceable or serviced only by Industrial
Scientific.
Содержание TANGO TX2
Страница 6: ......
Страница 16: ......
Страница 38: ...32 Illustration 7 1 Disassembled Tango TX2 and case top assembly Instrument Case top assembly...