Maintenance
iTrans Fixed Point Single/Dual Gas Monitor
7-2
INDUSTRIAL SCIENTIFIC
(P/N: 77023554)
NOTE:
If sensors and the surrounding environment must be washed down at any
time, cover the opening of the sensor housing to protect it from water or excess
moisture. Remove cover when wash down is complete. An optional splashguard
is available for continuous protection.
7.2. Sensor Replacement
Sensor replacement must be done by qualified personnel. To replace the sensor,
shut down power to the unit. Un-thread the sensor-housing cap from the sensor
housing. There is a set screw that secures the cap to the housing. Once the cap is
removed, remove the old sensor and sensor board. When installing the new
sensor/sensor board make sure you line up the notch in the board with the
alignment pin. After the new sensor is in place, screw the sensor cap back on to
the housing and secure the set screw.
Once the new sensor is in place and has time to settle out, it should be zeroed and
calibrated for accuracy.
7.3. Zero and Calibration
Zeroing and calibrating the instrument can be accomplished one of two ways.
These routines can be entered either from the keypad or non-intrusively using the
magnetic wand. See Chapter 5: Operation for step-by-step procedures for zeroing
and calibrating the
using the magnetic wand. Chapter 5 also contains
information on keypad zeroing and calibration.
NOTE:
When calibrating the BBIR sensor, the display will not show the actual
concentration of gas applied, it will display zero. Once the calibration is complete,
the display will switch from zero to the actual concentration of gas applied.
NOTE:
The typical zero time for the BBIR sensor is 3 minutes.
NOTE:
The typical calibration time for the BBIR sensor is 3 minutes.
NOTE:
After a calibration, the span reserve for the BBIR sensor is not available
to the end user.
Содержание iTrans 77023554
Страница 10: ...List of Tables iTrans Fixed Point Single Dual Gas Monitor x INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 16: ......
Страница 42: ...System Wiring iTrans Fixed Point Single Dual Gas Monitor 4 18 INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 86: ...Appendix D iTrans Fixed Point Single Dual Gas Monitor D 2 INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 88: ...Appendix E iTrans Fixed Point Single Dual Gas Monitor E 2 INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 94: ...Index iTrans Fixed Point Single Dual Gas Monitor vi INDUSTRIAL SCIENTIFIC P N 77023554 ...