iTrans Fixed Point Single/Dual Gas Monitor
Operation
(P/N: 77023554)
INDUSTRIAL SCIENTIFIC
5-7
NOTE:
Before the
will calibrate, the unit will enter the zeroing process.
Please make sure that you apply Zero Air to the instrument while it is zeroing.
Figure 5-10. Sample Zeroing
Display
The
will automatically zero
before calibration. Zeroing is designated
with a flashing “0” in the status bit. Once
zeroing is complete, the
will
automatically enter the calibration routine.
Calibration is designated with a flashing
“C” in the status bit.
After zeroing finishes, the
is
ready to calibrate. When the flashing “C”
appears on the display, apply calibration
gas. As the
responds to the gas,
the current reading will be displayed on the
top line of the LED display. To abort
calibration at any time, place the magnetic
wand over CH1.
NOTE:
When calibrating a BBIR sensor,
the display will not show the actual
concentration of gas applied, it will display
zero. Once the calibration is complete, the
display will switch from zero to the actual
concentration of gas applied.
Figure 5-11. Apply CalGas
Display
NOTE:
Placing the magnetic wand on CH1 during calibration, will cancel out of
calibration and return to the Programming Mode.
NOTE:
For combustible sensors the default value is set to 25% LEL Pentane.
NOTE:
See Appendix D for a complete list of factory default span gases.
NOTE:
Flow rate for calibration is 0.5 liter per minute (LPM) except for NH
3
,
ClO
2
, Cl
2
, NO
2
, SO
2
, and HCl which require 1.0 LPM.
NOTE:
Check and verify span setting before starting a calibration.
Содержание iTrans 77023554
Страница 10: ...List of Tables iTrans Fixed Point Single Dual Gas Monitor x INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 16: ......
Страница 42: ...System Wiring iTrans Fixed Point Single Dual Gas Monitor 4 18 INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 86: ...Appendix D iTrans Fixed Point Single Dual Gas Monitor D 2 INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 88: ...Appendix E iTrans Fixed Point Single Dual Gas Monitor E 2 INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 94: ...Index iTrans Fixed Point Single Dual Gas Monitor vi INDUSTRIAL SCIENTIFIC P N 77023554 ...