iTrans Fixed Point Single/Dual Gas Monitor
System Wiring
(P/N: 77023554)
INDUSTRIAL SCIENTIFIC
4-5
4.5. Sensor Wiring (J3)
Connect the
sensor wires (for on-board, remote or stand-alone) to the
appropriate wiring terminals as follows. Refer to Figure 4-2 and Figure 4-5.
24 V:
Red wire from sensor head
(Red wire on BBIR)
485B:
Yellow wire from sensor head
(White wire on BBIR)
485A:
Black wire from sensor head
(Green wire on BBIR)
GND:
Green wire from sensor head
(Black wire on BBIR)
NOTE:
Shielding from either the controller or remote sensors should be bonded
to the enclosure screw located inside the
.
NOTE:
The 24 V terminal supplies 24 VDC to the sensor for power. This
terminal should not be connected to the output of a 24 VDC power supply.
Figure 4-5. Sensor Connector J3 on the
NOTE:
For dual-sensor configurations, place both of the same colored wires in
the appropriate terminal block and firmly tighten.
NOTE:
Use #18 AWG shielded cable for remote sensors. Maximum distance is
200 meters (BBIR maximum distance is 300 feet with #18 AWG wire).
NOTE:
When wiring remote sensors to the
, “485 B” on J3 should be
connected to “B-” in the remote sensor enclosure, and “485 A” on J3 should be
connected to “A+” in the remote sensor enclosure.
Содержание iTrans 77023554
Страница 10: ...List of Tables iTrans Fixed Point Single Dual Gas Monitor x INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 16: ......
Страница 42: ...System Wiring iTrans Fixed Point Single Dual Gas Monitor 4 18 INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 86: ...Appendix D iTrans Fixed Point Single Dual Gas Monitor D 2 INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 88: ...Appendix E iTrans Fixed Point Single Dual Gas Monitor E 2 INDUSTRIAL SCIENTIFIC P N 77023554 ...
Страница 94: ...Index iTrans Fixed Point Single Dual Gas Monitor vi INDUSTRIAL SCIENTIFIC P N 77023554 ...