36
37
E. EMC Information (Declaration)
The BPBIO 320S is intended for use in the electromagnetic environment specified below. The customer or the
user of the BPBIO 320S should assure that it is used in such an environment.
Electromagnetic emissions
Emissions test
Compliance Electromagnetic environment
RF emissions CISPR 11
Group 1
The BPBIO 320S uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions CISPR 11
Class A
The BPBIO 320S is suitable for use in all establishments, including
domestic establishments and buildings directly connected to public
low-voltage power supply networks.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations
flicker emissions
IEC 61000-3-3
Complies
Electromagnetic immunity – for all ME equipment and ME systems
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT
for 0.5 cycle
40 % UT
(60 % dip in UTfor
5 cycles
70 % UT
(30 % dip in UT
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the BPBIO 320S requires
continued operation during power mains
interruptions, it is recommended the
BPBIO 320S be powered from an
uninterruptible power supply or battery.