E. Replacing Cuff Cover
The cuff cover of the BPBIO 320S should be replaced periodically for sanitary purposes. Follow the guide
below to replace the cuff cover.
1.
Remove the front cover of the BPBIO 320S by pushing downward. Loosen the screw with a screwdriver.
2.
Loosen the two screws on the rear of the BPBIO 320S with a screwdriver.
3.
Remove the rings on the front and rear of the BPBIO 320S by
pulling the arrow mark and pull the cuff cover to remove it.
4.
Place the seam of the new cuff cover down
as shown, and align it with the front and rear ring.
Screws
Rear ring
Front ring
Screws
Front Cover
•
Blood pressure results may be inaccurate if the seam is not face down.
Caution
•
Le résultat du test de la pression artérielle peut ne pas être correct si la ligne de couture n'est pas orientée
vers le bas.
Attention