Question
Cause
Measures
The device is not
turning on.
The fuse is blown.
Turn off the BPBIO 320S, and replace the fuse.
The fuse socket is located on the back of the device beneath
the power plug.
Pull the fuse out by using a flathead screwdriver. Replace
the fuse with the provided spare fuse. Additonal fuses can
be purchased if necessary.
Voice guide does not
work.
The volume knob on
the rear of the blood
pressure monitor
may be turned off.
Turn on the volume knob on the rear of the BPBIO 320S.
The sound sentting
has been set to ‘Beep
Guide’ instead of
‘Speaker’.
Refer to Setup on page 13 to change the sound settings.
TYPE: Fast-Acting
Rated current: 2.5AL
Rated voltage: 250V
Spare fuse
• Be sure to turn off the power when replacing the fuse.
• The correct rated fuse must be used. (250V, 2.5AL)
Warning