Environmental Equipment, Inc.
IMR EX610 User Manual
3632 Central Ave. St. Petersburg, FL 33711 – Phone: 727-328-2818 / 800-RING-IMR – Fax: 727-328-2826 – E-mail: [email protected] – Web: www.imrusa.com
THE PRIDE OF THE COMPANY – MADE IN THE USA
Page
15
of
27
TWA Alert
:
Slow and hi alarm sound
Flickering alarm lights
Vibration
①
Over range Alert:
Slow and hi alarm sound
Alarm LED flickering
Time fault:
The device will attempt to solve this problem automatically. If successful the device
will enter into the time setting after power on. The time can then be set.
If the device is unable to resolve the issue it will power off. Contact IMR.
Storage fault:
The device will attempt to solve this problem automatically. If successful the device
will enter into the calibration mode after power on. Refer to section 4.7 for the
procedure. If failed, the device will turn off automatically.
If the device is unable to resolve the issue it will power off. Contact IMR.
Low voltage alert:
Hi alarm sound once every 1 minute.
With very low-voltage the device can work for up to 15 minutes. Charge the
device in a non-hazardous area. Otherwise, it will turn off automatically.
Note:
①
Only available when vibration function is on.
②
Only available for Combustible Gas detector.
When the alarm is continuously sounding press
to stop audio alarm, and vibration. The
and
icon
on the screen will flicker indicating that the alarm status is currently suppressed.
4.6
Status Checking
In detection mode press
the device will then cycle through temperature, time, Max. Value, Min Value
②
,
STEL Value
①
, TWA Value
②
, etc.
Note:
①
Only available for toxic gas detector
②
Only available for Oxygen detector