5.
Data Uploading (If Equipped)
Connect the detector to a PC with a USB cable. Then open the data software.
Note:
This function will be available only if the product that was purchased is capable of communication with a
PC.
6.
Using and Replacement of sensors
It is important to be aware of the life span of equipped sensors. When they are beyond their expected life they
should be replaced. Approximately every half year calibration should be checked for accuracy in order to ensure
safety of the end user.
Sensor replacement should only be done by IMR or an authorized repair center. Only authorized parts should be
used. Contact IMR if any additional information is needed.
7.
Trouble Shooting Guidance
Trouble
Possible Reasons
Solutions
Unable to turn on the detector
Battery voltage is low
Charge the detector
Detector is defective
Contact the IMR
PCB fault
Contact the IMR
No response to gas
Warm up has not finished
Wait longer for it to finish
PCB fault
Contact the IMR
Gas response inaccurate
Sensor replacement overdue
Replace the sensor
Calibration overdue
Calibrate the detector
Date/Time is incorrect
Battery needs replacement
Replace battery and set
date/time
Electromagnetic interference
Set the date/time
Unable to set a new Zero calibration
Excessive sensor drift
Calibrate or replace the sensor
Display shows “-0”
Excessive sensor drift
Calibrate or replace the sensor
If the display shows
Sensor fault
Replace the sensor