The buzzer will beep. After 3 seconds. When the screen displays the following release
and the detector will
turn off.
Note: When the device is not in the detection interface. Press
again and again until the device exits to the
detection interface.
4.3
Bump Test
Perform a “Bump Test” before use each day to ensure that the detector is functioning properly.
Method:
After powering on the device. Expose the device to gas with a concentration that is higher than the H-
alarm level with standard test gas. If the device is working normally then the device can be used. If the response
on the screen is beyond the error range refer to section 4.8 to re-calibrate the device. If the device does not
respond or displays unexpected output, contact IMR.
4.4
Settings Menu
This menu includes the following:
Zero translation
Date and time
Vibration on / off
Keytone on/off
Communication mode
①
Password
Note:
①
only available when the purchased instrument is supplied with USB communication cable.