Page 30
Operation Manual
The screen content changes and shows the setup menu for white calibration.
Perform the white calibration with the glass plate closed by using
the “Calibrate GP button”.
Picture 14: Setup menu, white calibration
8.
Move the glass plate upwards with the “Lift 3” upper arrow button. While pressing
on the arrow up button, a login mask will be shown. It is necessary to login as Admin.
(user = admin password= scan2net) The glass plate opening angle should be in V-
mode.
Picture 15: White balance without glass plate
Warning: The glass plate opening angle is not limited! The glass plate can crack if it hits the
neck of the device.
9. Perform the wh
ite balance without glass plate with the “Calibrate” button