8
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment
conditions
Electrostatic
discharge(ESD)
IEC61000-4-2
±
8kV contact
±
15kV air
±
8kV contact
±
15kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with a
synthetic material, relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC61000-4-4
±2kV for power
supply lines
±1kV for
input/output lines
±2kV for power
supply lines
±1kV for
input/output lines
Main power level should be
that of a typical commercial
or hospital environment.
Surge
IEC61000-4-5
±1kV differential
mode
±2kV common
mode
±1kV differential
mode
±2kV common
mode
Main power level should be
that of a typical commercial
or hospital environment
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC61000-4-11
0% Ut
For 0.5/1/250/300
cycle
70% Ut
(30%dip ln Ut)
For25/30 cycle
0% Ut
For
0.5/1/250/300
cycle
70% Ut
(30%dip ln Ut)
For25/30 cycle
Main power level should be
that of a typical commercial
or hospital environment. If
the user of the MODEL MKC-
X800 requires continued
operation times when the
main power may be
interrupted, powering the
MODEL MKC-X800 with an
uninterruptible power supply
or battery is recommended.
Power frequency
(50/60 Hz)
magnetic field
IEC61000-4-8
30 A/m
30 A/m
The power frequency
magnetic field should meet
the characteristics of a
typical location such as a
commercial or hospital
environment.
NOTE: A.C. main voltage prior to application of the test level.
Содержание MKC-X800
Страница 1: ...User Manual Digital Process Compact UHD Color Camera Model MKC X800 ...
Страница 14: ...14 CCU 85 10 Troubleshooting 86 Common problems and solutions Start up 86 Error Message 89 ...
Страница 36: ...36 4 4 2 Contrast adjustment CONTRAST PEDESTAL FLARE GAMMA KNEE HDR ...
Страница 53: ...53 0 63 As the value increases the Black Stretch function becomes stronger ...
Страница 73: ...73 When a setting data is saved on a USB memory any past set data will be overwritten and lost ...
Страница 84: ...84 9 Appearance Camera Head ...
Страница 85: ...85 CCU ...