Guidance and manufacturer's declaration - electromagnetic emissions
The Model MKC-230HD is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model MKC-230HD should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic
RF emissions
㻌
CISPR 11
Group 1
The Model MKC-230HD uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
㻌
CISPR 11
Class A
The Model MKC-230HD is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage
fluctuations/flicker
emissions IEC61000-3-3
Complies
Immunity test
IEC 60601 test level
compliance level
Electromagnetic environment
guidance
Electrostatic
㻌
discharge(ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
㻌
±8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2kV for power
supply lines
±1kV for input/output
lines
±2kV for power
supply lines
±1kV for input/output
lines
Main power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC61000-4-5
±1kV differential
mode
±2kV common mode
±1kV differential
mode
±2kV common mode
Main power quality should be that
of a typical commercial or hospital
environment
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC61000-4-11
<5% Ut
(>95% dip In Ut)
for 0.5 cycle
40% Ut
(60% dip In Ut)
for 5 cycle
70% Ut
(30% dip In Ut)
for 25 cycle
<5% Ut
(>95% dip In Ut)
for 5 cycle
<5% Ut
(>95% dip In Ut)
for 0.5 cycle
40% Ut
(60% dip In Ut)
for 5 cycle
70% Ut
(30% dip In Ut)
for 25 cycle
<5% Ut
(>95% dip In Ut)
for 5 cycle
Main power quality should be that
of a typical commercial or hospital
environment. If the user of the
MODEL MKC-230HD requires
continued operation during power
mains interruptions, It is
recommended that the MODEL
MKC-230HD be powered from an
uninterruptible power supply or
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at characteristic of a
typical location In a typical
commercial or hospital
environment.
NOTE Ut is the a.c.mains voltage prior to application of the test level.