Subject to technical changes!
12
Technical data
The
IKA
®
MTS 2/4 digital is maintenace-free. It is subject only to the
natural wear and tear of components and their statistical failure rate.
When ordering spare parts, please give the manufacturing number
shown on the type plate, the machine type and the name of the spare
part.
Please send in equipment for repair only after it has been clea-
ned and is free from any materials which may constitute a health
hazard. Use only cleansing agents which have been appro-
ved by
IKA
®
to clean
IKA
®
devices. To remove use:
Dyes
isopropyl alcohol
Construction materials
water containing tenside / iso
propyl alcohol
Cosmetics
water containing tenside / iso
propyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside
For materials which are not listed, please request information from
IKA
®
. Wear the proper protective gloves during cleaning of the devices.
Electrical devices may not be placed in the cleansing agent for the pur-
pose of cleaning.
Before using another than the recommended method for cleaning or
decontamination, the user must ascertain with the manufacturer that
this method does not destroy the instrument.
Repair:
Please send in instrument for repair only after it has been
cleaned and is free from any materials which may constitute a
health hazard.
For repair, please request the “
Decontamination Certificate
” form
IKA
®
or use the download printout of it from
IKA
®
If you require servicing, return the instrument in its original packaging.
Storage packaging is not sufficient. Please also use suitable transport
packaging.
Maintenance and cleaning
Design voltage
or
VAC
VAC
230 ± 10%
115 ± 10%
Design frequency
Hz
50 / 60
Input power
W
45
Speed range
1/min 0 - 1100
(infinitely adju-
stable)
Speed setting
Turn dial on front side
Speed display
Scale 0 - 1100
Drive
Speed controlled
asynchronous motor
Agitation stroke
mm
3 Ø
Shaking motion
horizontal, circular
Time switch
min
1-99 / continuoues operation
Display
digital
Perm. duration of operation
%
100
Perm. ambient temperature
°C
+5 to +40
Perm. relative humidity
%
80
Protection class acc.
DIN EN 60529
IP 21
Protection class
I
Overvoltage category
II
Contamination level
2
Fuse
2x T4A 250 V
(
IKA
®
-Id.Nr. 28 205 00)
Operation at a terrestrial
m
max. 2000 above
altitude sea level
Max. load
4 microtiter plates
Dimensions (W x H x D)
mm
185 x 105 x 320
(without attachment)
Weight
kg
2.7
www.
GlobalScientificSupply
.com
GLOBAL SCIENTIFIC SUPPLY
The Laboratory Supply Company!