background image

13

TEMPERING

Control

Control Mode

Fluids

Control Parameters

Internal (int)

activated

-

-

-

-

-

-

-

-

-

-
-

-

-
-
-
-
-

-
-

activated

activated

activated

95 °C
5°C
100 °C
100 %

± 0.0 K

60.0, 5.0, 0.3, 3 s, +1.00 K, -1.00 k
1.0, 15.0, 0.0, 90 s, +1.00 K, -1.00 k
-

Automatic

Accurate

Internal

Internal

Internal (Kp, Ti, Td, Ts, Prop_Bp, Prop_Bn)

2 Point Calibration

2 Point Calibration

Fast

Manual

External (ext)

External (ext)

IΔ T (int - ext)I
Power Output
Calibration

Reset Calibration

Heating

Maximum

Minimum

External

External

External (Kp, Ti, Td, Ts, Prop_Bp, Prop_Bn)

3 Point Calibration

3 Point Calibration

Information (Kp, Ti, Td, Ts, Prop_Bp, Prop_Bn)

Limits

Temperature Sensor

Notification

Hysteresis
Sound

Infinite
Beep

ON

CF.EG28.N10.80.8
CF.EG39.N20.80.16 
CF.EG44.N25.80.19  
CF.EG48.N30.80.22
UF.Si.N30.150.10LV
HF.Si.20.200.50
HF.Si.20.250.50A
H

2

O (Water)

Customized 1
Customized 2

TIMER

Set
Display

00:00 [hh:mm]
-

GRAPH

Automatic
Manual

Axis Assignment

Set Temperature
Actual Internal Temp.
Actual External Temp.

X - Axis

Y - Axis

15 min
30 min
60 min
120 min

Axis Scaling

Maximum
Minimum

-
-

-

activated
activated
activated
activated

activated

95 ºC
5 ºC

PROGRAMS

Program 1
...
Program 10

-
-
-

OUTGAS

Speed
Interval

1000 rpm
10 s

DISPLAY

Main Screen

Pump
Timer

-

activated

Factory Settings

Menu

PUMP

Limits

Speed

Pressure

Maximum

Maximum

Minimum

3200 rpm
1000 rpm
0.4 bar

MODE

A

C

B

-
-

activated

SAFETY

Password
Safe Temp. Confirmation

000
-

SETTING

Languages

Display

Sound
Factory Settings
Communication
Information

English

Background

Key Tone

Tempering

Safety

Intro

Device Name

Pump

Software

Firemware Update Info

Firmware Update Info

Black
White

Deutsch
...

-
activated

Circulator

-
-
-

yes

yes

yes

yes

yes

yes

-

activaed

D

-

Menu structure:

Содержание ICC control

Страница 1: ...20000004067a IKA ICC control ICC control_092017...

Страница 2: ...ay 2 Mains switch 3 Power socket 4 Buoyage 5 Heater 6 Stand 7 Pt 100 Pt 1000 temperature sensor 8 Handle 9 Fan 10 RS 232 port 11 USB port 12 Clamp 13 Cable clip 14 External temperature sensor socket F...

Страница 3: ...ions 7 Setting up 7 Filling and draining 7 Fluid Standard information for IKA fluid 8 Operator panel and display 9 Commissioning 10 Setting the safety temperature 11 Operation 11 Working screen at the...

Страница 4: ...us situation which if not avoided can result in injury Indicates practices which if not avoided can result in equipment damage Reference to exposure to a hot surface Hot surface WARNING Declaration of...

Страница 5: ...nt measures that inhibit re or explosions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please contact IKA application support if...

Страница 6: ...erature of the heater can be very high After a power failure during operation the device may start automatically depending on operating mode Transport the device with care Do not transport or empty th...

Страница 7: ...otection under the buoyage 4 Delivery scope ICC control Mains cables Screwdriver use for safety circuit USB 2 0 cabel A micro B User guide A warranty card NOTICE Preparations Setting up Securely x the...

Страница 8: ...1 Classi cation HF Heating Fluid CF Cooling Fluid UF Universal Fluid 2 Chemical composition Si Silicone oil EG Ethylene Glycol 3 Temperature range Minimum temperature Maximum temperature N Negative T...

Страница 9: ...and open the menu options Con rm the menu settings H Pump button Increase the pump speed setting value Navigation and changing the setting in the menu I Pump button Decrease the pump speed setting va...

Страница 10: ...ump button I In standby status activate the heating function by press ing the Temp button D The device start heating func tion meanwhile the pump start to run In working status stop the pump function...

Страница 11: ...p rpm Safe Temp C Lock This symbol means that the function of the keys for con trolling the device are disabled The symbol no longer appears if the functions are en abled once again by pressing the ke...

Страница 12: ...you press the Menu button J the system skips directly back to the working screen If you press the Back button K the system skips back to the previous screen Menu Menu navigation Press the Menu button...

Страница 13: ...0 22 UF Si N30 150 10LV HF Si 20 200 50 HF Si 20 250 50A H2O Water Customized 1 Customized 2 TIMER Set Display 00 00 hh mm GRAPH Automatic Manual Axis Assignment Set Temperature Actual Internal Temp A...

Страница 14: ...time interval over which the control deviation is determined and the respective control variable dependent on Kp Ti and Td is calculated Ts must be adjusted to match the response characteristic total...

Страница 15: ...end Note At the end of the program all device functions are switched off Edit Edit change program parameters Seg No Segment number Ctrl Sensor int ext Determines whether control is through the interna...

Страница 16: ...erface Con guration The functions of the interface connections between the stirrer machine and the automation system are chosen from the signals speci ed in EIA standard RS 232 in ac cordance with DIN...

Страница 17: ...d the internal setting temperature if 0 internal control Read the external setting temperature if 1 external control IN_SP_3 Read the safety setting temperature IN_SP_4 Read the pump setting speed IN_...

Страница 18: ...not listed please request in formation from IKA application support Wear protective gloves while cleaning the device Electrical device may not be placed in the cleansing agent for the purpose of clea...

Страница 19: ...6 Pump off Heating off Too low liquid level Check the liquid level and buoyage Error 07 Pump off Heating off Too high voltage Check the mains power Error 08 Pump off Heating off Too low voltage Check...

Страница 20: ...eco 8 Plastic bath 8 litre IB eco 18 Plastic bath 18 litre IB pro 9 Stainless steel bath 9 litre IB pro 12 Stainless steel bath 12 litre IB pro 20 Stainless steel bath 20 litre Bridges and Covers BS I...

Страница 21: ...esolution K 0 1 Temperature display TFT LCD Temperature display resolution K 0 01 Classi cation according to DIN 12876 1 Class III FL suitable for ammable and non ammable uids Safety circuit adjustabl...

Страница 22: ...e liable for freight costs The warranty does not cover worn out parts nor does it apply to faults resulting from improper use insufficient care or maintenance not carried out in accordance with the in...

Отзывы: