7
Correct use
Use:
Use
ICC
(
I
mmersion
C
irculator
C
ompact)
control
for heat-
ing and circulating fluids.
Intended Use: Tabletop Device
Range of use (indoor use only):
- Laboratories
- Schools
- Pharmacies
- Universities
This device is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply
network that also supplies residential areas.
The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not
supplied or recommended by the
IKA
®.
- If the device is operated improperly or in contrary to the
IKA
®
specifications.
- If the device or the printed circuit board are modified by
third parties.
Unpacking
Unpacking:
- Please unpack the device carefully
- In the case of any damage a detailed report must be sent
immediately (post, rail or forwarder).
Transport safety:
Remove the transport protection
under the
buoyage
(
4
).
Delivery scope:
-
ICC control
- Mains cables
- Screwdriver (use for safety circuit)
- USB 2.0 cabel A – micro B
- User guide
- A warranty card.
NOTICE
Preparations
Setting up:
- Securely fix the immersion circulator on the bath.
- Place the device on an even, stable, clean, nonslip, dry
and fireproof surface.
- Keep at least 20 cm of open space at the front and rear
side.
- When a plastic bath is used, please ensure that the heater
does not contact the bath.
- The place for installation should be large enough and
provide sufficient air ventilation to ensure the room does
not warm up excessively because of the heat from de-
vice radiates to the environment.
- Do not set up the device in the immediate vicinity of
heat sources and do not expose to sun light.
Filling and draining:
- Connect the mains plug and turn on the device with
mains switch (
2
).
- The low level icon ( ) appears on the display.
- Fill fluid to the bath.
Note:
Pay attention to the fluid level information on
the display:
—
Low level
—
High level
For draining, please refer to the bath operating instruc-
tions.