22
SETTING
Languages:
The “Language“ menu option allows you to select the
desired language�
Display:
The ”Display“ menu option allows you to change the
background color and brightness of the working screen�
You can also activate the “Firmware Update Info“ screen
in this menu�
Sound:
The ”Sound” menu option allows you to activate/deacti-
vate the key tone�
Factory Settings:
Select the ”Factory settings” menu option by turning
and pressing the rotary/push knob� The system requests
confirmation to restore the factory settings� Pressing the
”OK” button resets all the system settings to the origi-
nal standard values set at dispatch from the factory (see
“Menu structure”)�
Communication:
The ”Device name” menu option allows you to edit the
device name� This can relatively identify the paired station�
The ”Bluetooth
®
” menu option allows you to activate/de-
activate the ”Bluetooth
®
” function� A check mark shows
that the option is activated�
Information:
The ”Information” menu option offers you an overview of
the most important system settings of the device�
SAFETY
Time out:
In the menu option “Time out”,you can determine a time
limit in the event of a communication breakdown be-
tween the station and the WiCo� The station continues
to work with the preset target values until the preset time
value has expired� Following that, the station runs using
the preset safety temperature and safety speed�
Note:
The initial time out is 30 seconds and you can de-
fine up to 60 minutes for this time limit�
Speed:
In the “Speed“ menu option, you can specify a appropri-
ate and safe speed for specified temperature�
Note:
The factory setting of the safe speed is 1000 rpm
and is activated after the time limit is set (see “Time out”)�
Temperature:
In the “Temperature“ menu option, you can specify a
temperature that is appropriate and safe for the circulat-
ing task�
Note:
The initial safe temperature is 30 ºC and is acti-
vated after the time limit is set (see “Time out”)�
Password:
In the “Password” menu, the menu settings can be locked
by a 3-digit password�
Safe Temp� Confirmation:
This menu can confirm (safety temperature of the station)
at start-up of the set “Safe Temp”� A check mark (√) indi-
cates that the function is activated�
OUTGAS
In the menu option, you are allowed to set the outgas
speed from 1000 rpm to 4600 rpm and outgas interval
from 10 seconds to 240 seconds�
This function can be used when filling external devices
such as laboratory reactors�
M3-Input:
Ext. Standby:
External standby input to stop the device functions Tem-
per and Pump�
ON:
Activate the ext� standby function� Device functions
are stopped in the event of an inverted initial state (ON)
at the input�
Default Status: Define the default status (OFF) of the input
as “Open” (high level) or “Closed” (low level)�