21
MODULES:
In “Modules” the multifunction connection (14, see Fig� 31)
outputs and inputs can be configured�
The outputs M1 and M2 can be controlled via “PRO-
GRAMS”�
M1-Valve:
ON:
Switche the external valve to the active state (inverted
initial state)�
Note:
In a started „Program“ the M1 segment settings
have higher priority�
Default Status:
Define the default state (OFF) of the external valve as
“Open” or “Closed”� It depends on the valve type (nor-
mally open or closed)�
Refill:
Use the external M1 valve to automatically check the level
when operating with water�
Observe “Default Status” setting�
Low level: Set the valve switch-on point (ON)�
High level: Set the valve switch-off point (OFF)�
M2 (ON/OFF): MODULES M2-Output switch
OFF:
M2-output switch in initial state�
ON:
M2-output switch in inverted initial state�
Note:
The initial state of the M2-Output switch is defined
in “MODULES M2-Output” in “Default Status” as “Open”
or “Closed”�
Cool: Activate/deactivate the cooling function�
OFF:
cooling function is deactivated�
ON:
cooling function is activated�
Delete:
Deletes the selection highlighted in yellow (seg-
ment or program)�
Insert:
Inserts a new program segment after the selected
segment�
Save:
Saves changes�
View:
Temperature-time overview displays for the program with
segments of the selected program�
Note:
If hysteresis is set as "Ctrl�Mode +/-x�xx K" for one
or more segments in the program, the duration of the
program cannot be determined�
Press and turn the knob (R) to display the segment details�
Cancel:
Cancels the process�
Note:
Once the program has been started, the program
no�, segment no� (active/total) and the remaining duration
of the segment or hysteresis are displayed in the graph�
Note:
Prior to using the “Refill” menu option, check the
function of the buoyage�
The “Refill” menu option has higher priority than the pro-
gram M1 segment settings�
Temperature (int):
Temperature-dependent control of the M1-valve�
By entering the boundaries of “Higher” and “Low”, the
temperature range is set in which the M1-valve can be
opened�
If you activate a rule “Int / ext> Set Temperature” is an
automatic control of the M1-valve dependence on the
target temperature�
By the “Hysteresis” value (set temperature hysteresis) the
control stability can be optimized�
Note:
int / ext stands according to the selected control
mode (internal (int) or external (ext)) for the actual tem-
perature to be controlled�
M2-Output:
Alarm:
Activate the alarm output (switch contact)�
Error:
In case of an error the “M2-Output — Switch” is
activated (ON, inverted initial state)�
Warning:
In case of a warning the “M2-Output —
Switch” is activated (ON, inverted initial state)
When “Warning” is activated, “Error” is simultaneously
activated (see section “Error code”)�
Default Status:
Defines the default status (OFF) of the
“M2-Output — Alarm” as “Open” or “Closed”� It de-
pends on the alarm type (normally open or closed)�
Switch:
Activate the switch output�
ON: Switche output into the active state (inverted initial
state)�
Note:
When the “Program” is started, the M1 segment
settings have higher priority�
Default Status:
Defines the initial state (OFF) of the switch
output as “Open” or “Closed”� It depends on the switch
type (normally open or closed)�
Temperature (int)
Temperature-dependent control of the M2-switch�
By entering the boundaries of “Higher” and “Low”, the
temperature range is set in which the M2-switch can be
opened�
If you activate a rule “Int / ext> Set Temperature” is an
automatic control of the M2-switch dependence on the
target temperature�
By the “Hysteresis” value (set temperature hysteresis) the
control stability can be optimized�
Note:
int / ext stands according to the selected control
mode (internal (int) or external (ext)) for the actual tem-
perature to be controlled�