NAVAIR 17—2OAX—737
4.7.2 Connect the signal generator calibrator sensor module input to the TI RF output connector.
4.7.3 Press the signal generator calibrator INSTR PRESET key (Blue key, AUTOMATIC OPERATION
key), and ensure that the signal generator calibrator is in the FREQ measurement mode and tunes to the TI
FREQUENCY MHz switches setting. Set the signal generator calibrator for AM modulation
measurements with the DETECTOR set to PEAK ±12.Activate the signal generator calibrator 3 kHz LP
FILTER.
4.7.4 Verify that the signal generator calibrator indicates between 28.8% and 31.2% AM modulation, and
record the reading.
4.7.5 Verify that the TI meter indicates within ±1.2%(AM depth) of the recorded value in step 4.7.4.
4.7.6 Set the TI meter function switch to 0—100, and then set the TI 30 Hz tone modulation control to the
fully cw position.
4.7.7 Verify that the signal generator calibrator indicates >60% AM modulation.
4.7.8 Set the TI 30 Hz modulation control to the UNCAL [fully ccw (0)] position.
4.7.9 Set the TI 1020 Hz tone modulation control to the fully cw position.
4.7.10 Activate the signal generator calibrator 50 Hz HP FILTER, and the 3 kHz LP FILTER. Verify that
the signal generator calibrator indicates >60% AM modulation.
4.7.11 Set the TI 1020 Hz tone modulation control to the CAL (UNCAL lamp off) position, and the
9960 Hz tone modulation control to the CAL (UNCAL lamp off) position.
4.7.12 Activate the signal generator calibrator 300 Hz HP FILTER, and the >20 kHz LP FILTER.
4.7.13 Verify that the signal generator calibrator indicates between 28.8% and 3 1.2% AM modulation.
4.7.14 Set the TI 9960 Hz tone modulation control to the fully cw position.
4.7.15 Verify that the signal generator calibrator indicates >60% AM modulation.
4.7.16 Set the TI 30 Hz tone modulation control to the CAL (UNCAL lamp off) position, and the 9960
Hz tone modulation control to the CAL (UNCAL lamp off) position.
4.7.17 Set the TI FREQUENCY MHz switches to 108.100. Set the TI G/S DDM switch to 90 Hz.
4.7.18 Deactivate the signal generator calibrator 300 Hz HP FILTER, and activate the 3 kHz
LP FILTER.
4.7.19 Verify that the signal generator calibrator
indicates between 19.2% and 20.8% AM modulation.
4.7.20 Set the TI G/S DDM switch to 150 Hz, and the TI meter function switch to 0—30.
4.7.21 Verify that the signal generator calibrator indicates between 19.2% and 20.8% AM modulation,
and record the reading.
4.7.22 Verify that the TI meter indicates within ±0.8%of the recorded value in step 4.7.21.
4.7.23 Set the TI meter Function switch to 0—100, the G/S DDM switch to 0, the LOC—G/S switch to G/S,
and the LOC DDM switch to 90 Hz.
4.7.24 Verify that the signal generator calibrator indicates between 38.4% and 41.6% AM modulation.
16