7
Quanti-Tray Sealer PLUS User’s Manual
Symptom
Cause and/or Recommended Action
Green light takes more than 10
minutes to illuminate.
Call IDEXX Technical Service at +1 800 321 0207 or your local distributor.
Green light goes off during
operation.
The temperature of the heated roller is below the optimal sealing range. Wait for the
green light to appear before continuing to run samples.
Sealer is on, but motor does not
start when a Quanti-Tray is inserted.
• Make sure the sealer has reached sealing temperature (green light is on).
• Make sure the Quanti-Tray is face down in the rubber insert, with the white side of
the tray facing up.
• Make sure the rubber insert is pushed firmly into the roller until the motor engages
it.
The motor starts when a
Quanti-Tray is inserted but does not
pull the tray through.
• Make sure the Quanti-Tray is correctly seated in the rubber insert and that the
insert is pushed firmly into the roller until the motor engages it.
• Clean the rubber insert. See “Cleaning and preventive maintenance” for
instructions.
• Make sure the insert is completely dry. Wipe it, if necessary.
Quanti-Tray and rubber insert are
partially stuck inside the sealer.
Press and hold the
Reverse
button to reverse the motor until the tray is completely
ejected from the input slot. After the rubber insert is removed, make sure that the insert
is dry. Wipe it, if necessary.
Note:
If the tray has gone all the way into the sealer, do
not use the Reverse button.Doing so can cause the tray to become lodged inside the
sealer. (See the next symptom.)
Quanti-Tray and rubber insert are
stuck inside the sealer and are not
accessible from the input slot.
If the end of the tray is inside the inlet, remove the input shelf and the lower door, and
then press the
Reverse
button.
CAUTION:
Do not touch the heated roller!
Sealer makes a loud hissing noise
when sealing, or liquid drips from
the bottom of the sealer. Green
status light may go out after trays
are filled.
The tray was overfilled. The sealer automatically removes excess sample from the tray
and discards it into the internal drip tray.
To prevent the problem:
• Do not fill trays with more than 100 mL of sample.
• Avoid excessive water on the surface of the trays.
To clean the sealer see “Cleaning and preventive maintenance” for instructions on
cleaning the inside of the sealer.
Important
: Do not remove the outer shell of the sealer or tilt the sealer if sample is
dripping from the unit. Wait until dripping stops and the sealer is cool. See “Cleaning
and preventive maintenance” for instructions.
Debris on roller is affecting the seal
quality.
Check the roller for debris, and clean the roller if needed. See “Cleaning and
preventive maintenance” for instructions.
Содержание Quanti-Tray Sealer PLUS
Страница 1: ...Quanti Tray Sealer PLUS User s Manual ...
Страница 12: ...12 Quanti Tray Sealer PLUS User s Manual ...
Страница 13: ...Quanti Tray Sealer PLUS Manuel de l utilisateur ...
Страница 24: ...24 Manuel de l utilisateur pour Quanti Tray Sealer PLUS ...
Страница 25: ...Quanti Tray Sealer PLUS Manual del usuario ...
Страница 36: ...36 Manual del usuario de Quanti Tray Sealer PLUS ...
Страница 37: ...Quanti Tray Sealer PLUS Benutzerhandbuch ...
Страница 48: ...48 Quanti Tray Sealer PLUS Benutzerhandbuch ...
Страница 49: ...定量盘封口机 PLUS 用户手册 ...
Страница 60: ...60 定量盘封口机 PLUS 用户手册 ...
Страница 61: ...Quanti Tray Sealer PLUS ユーザーマニュアル ...
Страница 72: ...72 Quanti Tray Sealer PLUS ユーザーマニュアル ...
Страница 73: ...Zgrzewarka Quanti Tray Sealer PLUS Instrukcja użytkownika ...
Страница 84: ...84 Instrukcja użytkownika zgrzewarki Quanti Tray Sealer PLUS ...
Страница 85: ...Quanti Tray Sealer PLUS Manual do Utilizador ...
Страница 96: ...96 Manual do Utilizador do Quanti Tray Sealer PLUS ...