Quanti-Tray Sealer PLUS User’s Manual
Troubleshooting
IMPORTANT
: If any of the following solutions fails to resolve a problem, contact IDEXX Technical Service at
+1 800 321 0207.
Symptom
Cause and/or Recommended Action
1
Internal Temperature Error
The internal temperature of the sealer is over the allowed temperature. The heater
turns off automatically so it can cool to a safe temperature.
Overheating can be caused by periods of extremely heavy use or by using the
sealer in a high-temperature environment. Wait until the green light appears before
continuing.
Overheating can also be caused by debris on the sensors. Check the sensors
for debris and clean if needed. See “Cleaning and preventative maintenance” for
instructions.
2
Temperature Read Error
The roller temperature cannot be accurately read. Power the sealer off and then on.
A temperature read error can be caused by debris on the sensors. Check the sensors
for debris and clean if needed. See “Cleaning and preventative maintenance” for
instructions.
3
Access Door Error
The upper access door was opened while the unit was turned on. If the error appears
while the door is closed, check that the upper access door and the lower access
panel are fully seated.
4
Voltage Error
The input voltage to the sealer is fluctuating outside the normal range. The problem
could be caused by faulty input power. Try another power source. If the problem
persists, call IDEXX Technical Service at the number above. (This error may flash for a
moment a few seconds after the sealer is powered off. This is normal.)
5
Roller Temperature Error
The temperature of the heated roller is out of the optimal sealing range. The problem
could be caused by sealing multiple cold samples one after the other, using a sealer
that was stored in a cold room, or the presence of hot fluids in the sealing area. Wait
for the green light to appear before continuing to run samples.
6
Temperature Stability Error
The heated roller temperature jumped between consecutive reads. Check the roller for
debris, and clean the roller if needed. See “Cleaning and preventative maintenance”
for instructions.
7
Temperature Error during
Seal
The temperature of the heated roller dropped below sealing temperature while sealing
a tray. After the sealing process is complete, the error disappears, but the power light
remains amber until the unit heats to the ready state. The problem could be caused
by continuous sealing of cold samples or by trays leaking onto the roller, causing it to
cool. Wait for the green light to appear before continuing to run samples.
Amber power light does not come
on.
•
Make sure the sealer is plugged into a working outlet with the correct voltage.
•
Make sure the power switch is on.
•
Check fuses and replace if necessary. See
“
Changing fuses.
”
Содержание Quanti-Tray Sealer PLUS
Страница 1: ...Quanti Tray Sealer PLUS User s Manual ...
Страница 12: ...12 Quanti Tray Sealer PLUS User s Manual ...
Страница 13: ...Quanti Tray Sealer PLUS Manuel de l utilisateur ...
Страница 24: ...24 Manuel de l utilisateur pour Quanti Tray Sealer PLUS ...
Страница 25: ...Quanti Tray Sealer PLUS Manual del usuario ...
Страница 36: ...36 Manual del usuario de Quanti Tray Sealer PLUS ...
Страница 37: ...Quanti Tray Sealer PLUS Benutzerhandbuch ...
Страница 48: ...48 Quanti Tray Sealer PLUS Benutzerhandbuch ...
Страница 49: ...定量盘封口机 PLUS 用户手册 ...
Страница 60: ...60 定量盘封口机 PLUS 用户手册 ...
Страница 61: ...Quanti Tray Sealer PLUS ユーザーマニュアル ...
Страница 72: ...72 Quanti Tray Sealer PLUS ユーザーマニュアル ...
Страница 73: ...Zgrzewarka Quanti Tray Sealer PLUS Instrukcja użytkownika ...
Страница 84: ...84 Instrukcja użytkownika zgrzewarki Quanti Tray Sealer PLUS ...
Страница 85: ...Quanti Tray Sealer PLUS Manual do Utilizador ...
Страница 96: ...96 Manual do Utilizador do Quanti Tray Sealer PLUS ...