IDEXX EquiView

®

 Digital Radiography System Operator’s Guide 

21

Troubleshooting

This table identifies possible problems and procedures to fix them:

Problem

Potential cause

Troubleshooting procedure

EquiView Digital 
Radiography System 
does not turn on

•  No power
•  Faulty power switch on 

the computer

•  Check the power cord, plugs, and sockets
•  Check the fuses
•  Contact a service technician

Computer does not 
identify the digital 
X-ray plate 

•  No cable connection 

between the plate and 
the computer

•  Software not started

•  Check the cable connection from the plate 

to the computer and, if necessary, make the 
connection

•  Start the EquiView software and wait until the 

connection has been made

Starting the EquiView 
software produces 
“Error initializing 
device. Could not 
connect to Meteor II 
CL hardware...”

•  This message 

indicates that the 
EquiView software 
cannot communicate 
with the Matrox video 
card.

•  Make sure the plate cable is correctly 

connected to the computer. Try to run the 
EquiView software again. 

•  If the same message appears, try the 

following:

  Choose Start> System> Hardware and then 

click Device Manager. You should see a list of 
devices, including the Matrox Meteor II. The 
associated icon should be green with no other 
symbols. Click the icon and click Properties. 
The device should be enabled. If the device is 
disabled, click Enable. If this does not solve the 
problem, try restarting the computer by turning 
it off and waiting a few minutes before restarting. 
If you still get the error, contact IDEXX Technical 
Support.

The EquiView 
software opens but 
new X rays do not 
appear or an image 
with a broad white 
region appears.

•  Error in 

synchronization or 
error in connection to 
trigger synchronization 
unit.

•  Check synchronization. Is the trigger 

synchronization unit connected correctly? 
Check all connections, and check and/or 
replace batteries.

Image is 
underexposed or 
overexposed.

•  Wrong exposure 

factors 

 
 

•  Presets for first view 

images not optimal

•  Check that exposure factors on your X-ray 

source are set correctly. For example, 80 
kVp and an mAs of 1.8 (0.12 seconds on the 
MinXray) should give a good image of most 
leg joints in the horse.

•  It is possible to adjust the presets in 

the system that will globally change the 
appearance of the first view images (the 
view portrayed on-screen before any manual 
manipulation takes place). Your system was 
adjusted for your preference at installation. If 
your images are not diagnostic in the first view, 
contact IDEXX Technical Support.

NOTE:

 Error messages without titles generally emanate from the Microsoft Windows

®

 XP

®

 operating 

system software. Contact IDEXX Technical Support for assistance.

Содержание EquiView

Страница 1: ...IDEXX EquiView Digital Radiography System Operator s Guide...

Страница 2: ......

Страница 3: ...or industrial property rights covering this document or subject matter in this document The furnishing of this document does not give a license to these property rights except as expressly provided i...

Страница 4: ......

Страница 5: ...ing Images 12 Manipulating Images 14 Dynamic Views 14 Printing Images 17 Printer Settings 17 Preview Mode 17 Exporting Images and Creating CDs 18 Exporting Images 18 Creating a Patient CD 18 Viewing t...

Страница 6: ......

Страница 7: ...ng processing and archiving digital X ray images Computer The portable Microsoft Windows compatible computer runs the EquiView software which lets you view manage and archive the image files created f...

Страница 8: ...ribed in the Accessories section and if a hazard free combination with these devices components or assemblies cannot be deduced from the technical data you must consult the respective manufacturer to...

Страница 9: ...s a risk of damage Wait until the EquiView Digital Radiography System has warmed up to the ambient temperature and condensation has completely dried This may take several hours The initial installatio...

Страница 10: ...tion use the trigger cable to connect the switch box s detector socket to the computer 5 Turn the switch box to the on position 6 Connect the power cord to the computer and then to an appropriate elec...

Страница 11: ...tware indicated on the software disc 3 On the initial screen of the software there are two options The Take Images option lets you select or add a patient and then take X rays The View Images option l...

Страница 12: ...configured so that horse names are not displayed until a specified number of characters have been typed 2 When you locate the desired horse record click the name and then click the Continue button to...

Страница 13: ...of the breed to quickly access a particular breed in the list This is an optional field you can accept the default value Unknown 8 In the Age field enter the horse s age or use the up or down arrows...

Страница 14: ...t likely error that can occur is underexposure in which only a part of the dynamic range of the digital X ray plate is used In this case especially bony structures appear significantly less structured...

Страница 15: ...quipment TIP Use high kV and low exposure time Dose increases proportionally with the kV but to second order with time TIP Using old generators that do not have full wave rectification and have exposu...

Страница 16: ...n with that view You can skip image views if necessary or desired by selecting the check box next to that particular image Reshooting an Image To reshoot an image click the view name it turns blue the...

Страница 17: ...age taken while in the Take X Ray mode However you can display all of the images for the particular view when you are in the View Images mode Reviewing Images You can double click any image view name...

Страница 18: ...s display in their corresponding sections Search by Owner Click the owner s name in the Owners Found section This displays all horses for that owner in the Horses Found section To view the images of a...

Страница 19: ...navigation bar at the bottom to select individual views TIP At the lower left corner of each image is a toolbar with the following tools View single and multiple images Measure distances and angles S...

Страница 20: ...tings for the images and make the diagnosis Dynamic Views The mathematical treatment of the raw data from the plate is a very complex combination of functions known as an algorithm IDEXX has developed...

Страница 21: ...ossible osteophyte or enthesiophyte formation around the joint The dynamic range of the digital X ray plate and the EquiView software allow all these details to be seen in one image To adjust the Wind...

Страница 22: ...ld the Z key on the keyboard and then scroll with the mouse Or right click and scroll up Rotate To rotate an image counterclockwise press and hold the A key on the keyboard To rotate the image clockwi...

Страница 23: ...commend printing the images in Landscape as this allows the image to fill most of the paper size 2 Select the printer and choose the paper size 3 Click Print NOTE The EquiView Digital Radiography Syst...

Страница 24: ...ee types of images to indicate how you want the images to be exported 4 Select the location where you want to save the exported images NOTE The default location for exported images is the Images to Ex...

Страница 25: ...stalled on your computer If the software is not already installed follow the instructions below To install the Microsoft NET Framework software 1 The patient CD is equipped with the NET Framework soft...

Страница 26: ...ecting Disable Auto Protect You will be asked to choose a duration select Until system restart 3 Stop the SQL Server Service by right clicking the SQL Server icon and selecting MSSQLServer Stop When y...

Страница 27: ...r symbols Click the icon and click Properties The device should be enabled If the device is disabled click Enable If this does not solve the problem try restarting the computer by turning it off and w...

Страница 28: ...osive solvent or gaseous disinfectants If you use disinfectants that form an explosive gas mixture they must evaporate first before turning on the device again Turn off the power before disinfecting t...

Страница 29: ...m IDEXX Technical Support The equipment can only be used again after repair Operation with defective components can lead to an increased safety risk Interval Scope Method Daily Damaged parts inscripti...

Страница 30: ...ced into the case between taking shots A combined heating and cooling element Peltier element is installed in the transport case and insulation ensures that the temperature within the case changes ver...

Страница 31: ...te After connecting the power supply it is absolutely necessary that the temperature control of the case be switched on using the switch and the case closed This ensures that the valuable plate has a...

Страница 32: ...rtable unit Active area 20 cm x 25 cm 8 x 10 Electrical Data Computer Power supply 100V 240V AC 10 50 60 Hz Plate case Power supply 100V 240V AC 10 50 60 Hz 75 Wh power consumption or 12V DC using the...

Страница 33: ...nsation can form in the device Operate the device only under the above mentioned environmental conditions Computer Specifications CPU Intel Pentium 4 processor 2 4 GHz 512 KB cache 400 MHz system bus...

Страница 34: ...e your finger left to right and up and down on the trackpad to adjust the settings for window and level respectively Zoom Right click and hold and then move your finger up and down on the trackpad to...

Страница 35: ...1 1 5 20 24 DP 66 0 12 1 8 20 24 Carpus All views 66 0 1 1 5 20 24 Skyline 68 0 14 2 1 12 16 Tarsus Lat Obl 70 0 1 1 5 20 24 DP 70 0 12 1 8 20 24 Stifle Lat Obl 70 0 2 3 20 24 CC 70 0 4 6 close as po...

Страница 36: ...One IDEXX Drive Westbrook Maine 04092 USA 06 14899 00...

Отзывы: